background image

 

User‟s Manual  

for Victor 88E Digital Multimeter© 

 

                                  

 

Distributed and Copyrighted by 

www.FiredragonEnt.com

 

Содержание Victor 88E

Страница 1: ...1 User s Manual for Victor 88E Digital Multimeter Distributed and Copyrighted by www FiredragonEnt com...

Страница 2: ...2...

Страница 3: ...ACV DCA ACA resistance capacitance frequency temperature duty cycle and diodes and acts as a continuity performance tester The meter can provide functions including analog bar and unit symbol display...

Страница 4: ...k the connection and insulation of test leads to avoid electric shock 2 Do not input a limited voltage higher than DC 1000V or AC 750V when measuring 3 When measuring voltage higher than DC 60V AC 40V...

Страница 5: ...symbol display Measurement method double integral A D conversion Sampling rate approximately 3 times sec Over range display OL displayed in the highest digit Low battery display Working environment 32...

Страница 6: ...ure Probe Duracell batteries and P25 hard case The Victor 88E C comes with Standard Package L36 Test Leads FSG1 Flame Safeguard Adapter GA1 and GA2 Gas Burner Thermocouple Adapters PT1 Probe Pins and...

Страница 7: ...1 0 4 1V Input impedance 400mV Range 10M other range is 40M Overload protection 1000V DC or 750V AC AC Voltage ACV Range Accuracy Resolution 400mV 1 5 6 0 1mV 4V 0 8 6 1mV 40V 10mV 400V 100mV 750V 1 0...

Страница 8: ...seconds Overload protection 400mA 250V and 20A 250V fast acting fuse AC Current ACA Range Accuracy Resolution 400uA 1 5 5 0 1 A 4000uA 1 A 40mA 10 A 400mA 100 A 20A 2 0 15d 10mA MAX measurement voltag...

Страница 9: ...protection 250V DC or AC peak value NOTE At a range of less than 400 short circuit the test leads to measure the wire resistance and then subtract it from the real measurement Capacitance C Range Acc...

Страница 10: ...value Diode and continuity performance test Range Value displayed Testing condition Forward voltage drop of diode Forward DCA is approx 0 5mA the backward voltage is approx 1 5V Buzzer makes a long so...

Страница 11: ...nge Accuracy Resolution 40 C 1000 C 0 C 1 0 4 0 C 400 C 1 0 4 400C 1 5 15 1 C 0 F 1832 F 750 F 0 8 5 750 F 1 5 15 1 F Sensor TP01 K type thermocouple CAUTION DO NOT INPUT VOLTAGE WHEN OPERATING IN THI...

Страница 12: ...it will return to auto range mode 2 3 HOLD key press it the presently measured value is held on LCD and HOLD symbol displays Press it again HOLD symbol disappears and the meter exits the holding mode...

Страница 13: ...suring the Frequency it will switch frequency duty cycle 1 99 3 Selector Knob Selects measuring function and range 4 TERMINAL Temperature and V Hz Voltage Resistance Frequency Capacitor and Temperatur...

Страница 14: ...l across the electric circuit under test LCD displays polarity and voltage under test connected by the red test lead Note 1 Users should first set the select knob to the correct range if users have no...

Страница 15: ...RANGE key will display AC mV range 4 Connect the leads in parallel across the electric circuit under test LCD displays voltage by the test lead Note 1 Users should first set the selector knob to the...

Страница 16: ...n connect the leads in series into the electric circuit under test LCD displays current under test Note Users should first set the selector knob to the highest range if users have no idea about the ra...

Страница 17: ...the highest range if users have no idea about the range of current under test and then select the proper range based on displaying value 2 If the LCD displays OL it means the current is over range Now...

Страница 18: ...This is normal for high resistance measuring 2 When the input lead is in an open circuit overload displays OL 3 When measuring in line resistors be sure that the power is off and all capacitors are c...

Страница 19: ...acitors The capacitance must be complete discharged before testing to protect the meter A reading over the 400uF range will delay 15 seconds Frequency Measurements 1 Insert the black lest lead into th...

Страница 20: ...s not guaranteed In an RFI noise environment you should use shielded cable to measure a low signal When measuring high voltage circuits no part of your body should touch the high voltage circuit Never...

Страница 21: ...red test connected to the diode cathode polarity and the black test lead to the diode positive polarity the LCD will display OL 4 This will complete diode testing including forward and backward measu...

Страница 22: ...the object to be tested Read the temperature from the screen and the data is in Centigrade 3 Press the SELECT key to select Fahrenheit the data is in degrees Fahrenheit Note 1 When the input terminal...

Страница 23: ...er powers off Press any key to restart the power Meter Maintenance The 88E is a high accuracy precision instrument please do not attempt to modify any of the internal circuits 1 Keep the meter away fr...

Страница 24: ...Never measure a voltage with the meter selector knob in the current resistance temperature capacitance or diode and continuity range 3 Never use the meter without a battery installed and the back case...

Страница 25: ...fications are subject to change without notice We hereby will not be responsible for any accident and damage caused by improper operation The functions stated in this User Manual cannot be exceeded by...

Страница 26: ...ning To prevent electric shock fire and personal injury carefully read all safety information before attempting to operate the current Clamp and follow these procedures Do not use the clamp on circuit...

Страница 27: ...F 30 C to 40 C 75 RH 104 F to 122 F 40 C to 50 C 45 RH Storage conditions 22 F to 140 F 30 C to 60 C 90 Relative Humidity Altitude Operate at less than 9800 feet or 3000 meters Weight 4 5 ounces Dimen...

Страница 28: ...t to select AC push SELECT button on 88E Position the Klamp jaw perpendicular to and centered around the conductor to be measured Read the measured value from the 88E LCD display Example for using mul...

Страница 29: ...ll powerburner PSC motor multimeter reads 1 30 mV actual reading is 1 3 amps Figure 2 Description of Amp Klamp 88CP9 1 Current sensing jaw 2 Klamp trigger to open 3 Safety barrier 4 Voltage output lea...

Страница 30: ...30...

Отзывы: