The WaveShaper can be controlled using the web browser on a hand-held device such as a tablet or phone. However, performance
has been optimised for use with a mouse and keyboard and not all functions may operate correctly when used with a touchscreen
device.
2.4
Can't find the WaveShaper?
If the WaveShaper connected to the network does not appear in the WaveShaper Connect window, then try these steps:
Ensure that the WaveShaper is connected to the LAN
before
powering-on the instrument. The network will only assign
an IP address to the WaveShaper if it is connected to the network before powering on. The default WaveShaper IP
address may cause some issues in some network configurations.
If the WaveShaper still does not appear, it could be due to changes made to the WaveShaper IP settings. The
WaveShaper IP setting can be reset to the default mode with the reset switch. The default mode is
"DHCP then Fixed
IP".
While the WaveShaper is running, press and hold the reset switch at the back of the instrument for at least
5
seconds
.
Let go of the reset switch and afterwards power cycle the WaveShaper.
The WaveShaper should then be discoverable in the WaveShaper Connect window.
For further detailed technical help, please visit the Finisar KnowledgeBase website.
If there are still connection issues, then please contact us by email: waveshaper@finisar.com
2.5
WaveShaper Mac Address
For both the Module and Benchtop, the MAC addresses of the USB and Ethernet interfaces are printed on the Serial Number
Label.
WaveShaper A Series User Manual
8