background image

4

Heel Bumper Installation 

The foot is supplied with two heel 
bumpers: 40A durometer is standard 
and pre-installed. If desired, the “Firm” 
bumper (60A) can be installed to 
stiffen the heel performance.

1. 

Pull the pylon and foot plate 
apart. Insert a piece of plastic to 
hold parts open. 

2. 

 Remove the Heel Bumper by pulling it up and away from the heel lock.

3. 

 Install the new bumper and ensure the Lock Pins engage the Heel 
Bumper properly.

4. 

 When the desired bumper density is determined, remove the bumper 
and apply Superglue to the Heel lock and reinstall the bumper

Static Alignment - Sagittal Plane

Before aligning, the initial heel height should be established. Using the 
adapter surface for reference, the AllPro employs a 2° posterior lean 
(Figure 1) with a 10 mm (3/8 in.) heel block to preload the anterior keel. When 
the patient is weight bearing, the adapter surface should settle to a neutral 
or level position. 

Transtibial Frontal Plane Alignment 

A plum line from the bisection of the socket 
at the proximal brim in the frontal and 
sagittal plane should bisect the ankle pyramid 
(Figure 1). The foot may be slightly inset 1 – 12 
mm depending on the limb length. Most 
runners prefer a wider base of support with 
the foot slightly lateral to the distal bisection 
7 – 13 mm. The longitudinal axis of the foot will 
be externally rotated approximately 5 – 8° by 
aligning the medial border of the foot with the 
line of progression (Figure 2).

Figure 3

Apply Glue

Pin Lock

Heel Lock

Heel 

Bumper

Figure 1

Содержание AllPro Foot

Страница 1: ...AllPro Foot Product Manual...

Страница 2: ...ual on page 7 before use of the foot These instructions should be read prior to fitting and followed to ensure the proper integration of the AllPro foot into the patient s prosthetic system The foot s...

Страница 3: ...use the epoxy adhesive provided or equivalent to secure the pyramid assembly Fabtech systems PLUSeries adhesives are acceptable 3 Apply adhesive to the exterior faces of the RPI adapter and push into...

Страница 4: ...be established Using the adapter surface for reference the AllPro employs a 2 posterior lean Figure 1 with a 10 mm 3 8 in heel block to preload the anterior keel When the patient is weight bearing th...

Страница 5: ...ing this process is essential Adjustments of the plantar dorsiflexion angles will help the patient achieve a smooth transition from heel to toe Check for smoothness of gait and ground contact througho...

Страница 6: ...hell onto the foot until the toe is in position Rotate the foot side to side to allow the foot shell to slide onto the heel Push foot shell over the heel or if necessary insert shoehorn into foot shel...

Страница 7: ...my AllPro wet The AllPro is designed to be maintenance free The foot is water resistant however if the foot is submerged in water the foot and foot shell should be rinsed with fresh water and dried im...

Страница 8: ...follow the weight guidelines and or overload conditions caused by the patient such as heavy lifting high impact sports or abusive activities that would otherwise damage the natural limb may void the w...