Installation

Attention: 

Deviating from the installation instructions or modifying the foot 

in any way will void any product warranty and could lead to product failure 
and injury to the patient.

Reversible Pyramid Insert (RPI) 

The RPI is set at 35% of foot length, 
except for AK models which are set at 
30% of foot length. If the BK user would 
benefit from more dynamic response, 
the RPI adapter can be flipped 180°. See 
instructions below for final assembly. 

1. 

Walk patient in both positions to 
determine the desired pyramid 
position.

2. 

Before final delivery of the 
prosthesis, use the epoxy adhesive 
provided or equivalent to secure 
the pyramid assembly. Fabtech 
sPLUSeries adhesives are 
acceptable.

3. 

Apply adhesive to the exterior 
faces of the RPI adapter and push 
into desired position.

4. 

Apply adhesive to the bushing shaft and insert into hole of RPI adapter.

5. 

Apply adhesive to the top surface of RPI insert and bushing shaft. Then 
install pyramid dome and ensure pin location lines up with the insert 
pin hole.

6. 

Apply adhesive to the top of the bolt threads and insert bolt through 
bushing.

7. 

Use 6 mm hex wrench to torque bolt to 30 N m.

8. 

Wipe away any excess adhesive with towel wetted with alcohol.

Pyramid 

Dome

Pyramid 

Insert

Bolt

Bushing

30% Position

(Increased 

Dynamics)

Pyramid 

Dome

Pyramid 

Insert

Bolt

Bushing

35% Position 

(Standard)

Содержание AllPro Foot

Страница 1: ...AllPro Foot Product Manual...

Страница 2: ...ual on page 7 before use of the foot These instructions should be read prior to fitting and followed to ensure the proper integration of the AllPro foot into the patient s prosthetic system The foot s...

Страница 3: ...use the epoxy adhesive provided or equivalent to secure the pyramid assembly Fabtech systems PLUSeries adhesives are acceptable 3 Apply adhesive to the exterior faces of the RPI adapter and push into...

Страница 4: ...be established Using the adapter surface for reference the AllPro employs a 2 posterior lean Figure 1 with a 10 mm 3 8 in heel block to preload the anterior keel When the patient is weight bearing th...

Страница 5: ...ing this process is essential Adjustments of the plantar dorsiflexion angles will help the patient achieve a smooth transition from heel to toe Check for smoothness of gait and ground contact througho...

Страница 6: ...hell onto the foot until the toe is in position Rotate the foot side to side to allow the foot shell to slide onto the heel Push foot shell over the heel or if necessary insert shoehorn into foot shel...

Страница 7: ...my AllPro wet The AllPro is designed to be maintenance free The foot is water resistant however if the foot is submerged in water the foot and foot shell should be rinsed with fresh water and dried im...

Страница 8: ...follow the weight guidelines and or overload conditions caused by the patient such as heavy lifting high impact sports or abusive activities that would otherwise damage the natural limb may void the w...

Отзывы: