Quick Start

1.   Press 

 for 1 second to turn on your SDMN6.

2.  If necessary, press ZERO for 1 second to set the 

ambient pressure of P1 and P2. 

3. P1 will be displayed in inches of water column 

(inWC.) P1-P2 will be displayed on lower line.

4. See sections below for details on testing pres-

sure switches.

                                   Scan for Video

Certifications

 

   C-Tick (N22675)

   CE 

  

   WEEE

    

     

RoHS Compliant

Description

Correct pressure switch operation is integral 

to the safety of a furnace environment. 

The SDMN6 is a dual-port manometer that  

also accurately tests pressure switches by 

simulating a draft with an internal pump.

Adjust the pump's speed to create a negative 

pressure (vacuum). The red LED indicates when 

the switch closes. View the closing/opening 

pressures directly on the LCD.

The dual-port manometer measures ±60 

inches of water column, high enough for gas 

pressures. 

Check building pressurization with static 

pressure resolution of 0.01"WC. View P1-P2 

in the bottom line of the display for checking 

pressure differentials. 

Use the static pressure probes to check for 

a pressure drop accross two points in a duct. 

This is useful for evaluating a blower or finding 

airflow restrictions in the ductwork.

Display

Controls

 

  

Hold for one second to toggle power.  

  

 

Press to toggle backlight.

  Hold for one second to zero  

  

 

pressure of P1 and P2.

  Select P1 or P2 on top display.

   

(Hold while powering ON to disable Auto-off.)

  Select mbar, inWC, mmWC, or psi.

   

(Hold to check % battery remaining.)

Fieldpiece

Dual Port  

Manometer  

and Pressure  
Switch Tester

OPERATOR'S 

MANUAL 

Model 

SDMN6

   

HOLD the displayed measurements. 

  In TEST mode:  press to activate auto-capture mode.  

  

The pressure reading will automatically HOLD  

  

when the pressure switch closes or opens. 

  Enter pressure switch TEST mode and start the  

  

 

pump (allow warm up to finish). Press again  

  

 

to stop the pump and return to P1/P2 mode.

  In TEST mode, increase pump speed (InC) to increase  

  

the induced vacuum. (Holding increases faster.)

  Decrease pump speed (dEC) to decrease the induced  

  

vacuum. (Holding decreases faster.)

Static Pressure 

1. ZERO your SDMN6 while at ambient pressure 

with any hoses/probes attached. This will zero 

both P1 and P2. For measurements less than 

2 inWC, take reading within 1 minute after 

zeroing for best accuracy.

2.  Use one hose (P1 or P2) to get the gauge pres-

sure relative to the ambient or ZERO pressure.  

3.  Connect both hoses if you want to see rela-

tive pressure, P1 - P2. This is common when 

evaluating equipment such as a blower. 

NOTE:  Always use the static pressure probes if 

checking static pressure of an airflow stream.

NOTE:  The red arrows on the probes should 

point toward the airflow stream.

>> 

AIRFLOW DIRECTION 

>>

        

 

 

4.  You can press P1 / P2 to toggle the top display 

between P1 and P2.  Note:  P1 - P2 is shown 

in the bottom line of the LCD.

5.  Press UNITS to select:  inches of water column 

(inWC), millimeters of water column (mmWC), 

millibar (mBar), or pounds per square inch 

(psi). 

6. Your SDMN6 will turn off automatically after 

15 minutes of inactivity if APO is shown. 

To disable auto-power-off (APO) hold P1/P2 

while turning on your SDMN6.  

7. If you are in an environment where the tem-

perature is noticeably changing while you are 

taking your reading, it is advised that you dis-

connect the meter from the hoses and ZERO 

it relative to ambient before each reading.

Gas Pressure 

1. See manufacturer's specification for target 

inlet and outlet pressures.

2.  Shut off main gas supply to furnace.

3.  Shut off power to furnace equipment.

4.  Zero your SDMN6 while at ambient pressure 

with hoses attached. 

5.  Remove outlet pressure tap screw, and insert 

the brass screw fitting of the P1 hose into the 

outlet pressure tap of the gas regulator.

NOTE: Some pressure taps have a 5/16" boss 

instead of an insert screw. In this case, remove 

the brass screw fitting from the P1 hose and 

connect the included 5/16" adapter tube.  

Loosen the tap screw about 1 revolution, and 

slide the tube over the boss (see  illustration 

at left). Make sure the tube is over the boss 

enough to prevent leakage.

6.  Put furnace into operation (i.e. turn on gas 

and power to the furnace, call for heating, and 

have furnace ignite as if running it in normal 

operation).

7. P1 will show the pressure coming out of the 

regulator.

8.  If you suspect high or low inlet pressure into 

the regulator, connect P1 to inlet tap and P2 

to outlet tap to see the pressure drop across 

the regulator.

Smaller end (3/16”) of the adapter 

connects to SDMN6 tubing. Remove 

the installed brass screw fitting 

from tube so the adapter can be 

inserted.

Adapter

(5/16” to 3/16”)

Adapter Tube

(5/16”)

Outlet Pressure 

Tap Screw

Inlet Pressure 

Tap Screw

Outlet 

Pressure Tap 

Inlet Pressure Tap

Specifications

Operating environment:

 32°F to 118°F (0°C to 48°C) at <80% RH

Storage temperature:

 -4°F to 140°F (-20°C to 60°C), 0 to 80% RH 

(with battery removed)

Power:

 9V, NEDA 1604A, IEC 6LR61 9V alkaline battery.

Battery life (alkaline): 

200 hours (manometer only, backlight off),  

12 hours (pressure switch testing @ 27% pump speed, BL off, 

approx -2.5"WC induced pressure w/o pump bleed accessory)

Auto power off (APO):

 15 minutes

Low battery indication:   

Units of Measure:

 inWC, mmWC, mbar, psi

Resolution:

 0.01 inWC 

Overrange:  

“OL” or “-OL” is displayed

Max Pressure:  

900inWC (17.4psi) will damage sensors

Compatible media:  

Dry, non-corrosive gases

Pressure ports:

 2 connectors (P1, P2)  

for flexible tubing (4.5mm to 8mm ID)

Pump port:

 1 connector (PUMP)  

for 5mm (~3/16 inch) ID flexible tubing

Lead jacks:

 2 banana plug jacks for testing open/close switch status 

(continuity between terminals)

Accuracy:

 Stated accuracy at 0 to 50°C (32 to122°F): ±1.5% FS

Accuracy and ranges:

  inWC: ±0.02 on 0.00 to ±2.00 (±1.5% FS on 2.00 to ±60.0);

  mmWC: ±0.5 on 0.00 to ±51.0 (±1.5% FS on 51.0 to ±1500);

  mbar: ±0.05 on 0.00 to ±5.00 (±1.5% FS on 5.00 to ±150.0);

  psi: ±0.001 on 0.000 to ±0.07 (±1.5% FS on 0.07 to ±2.000)

inWC

PUMP

P1

P2

Gas 

Manifold

0.05

inWC

Blower

-0.03

inWC

inWC

PUMP

P1

P2

Get the free mobile app at

http://gettag.mobi

Blinks When 

Pump is ON

Leads Detect 

Continuity

Auto Power Off 

(APO) Enabled

Real Time Pressure 

Measurement of P1 or P2

(mbar, inWC,mmWC, or psi)

Solid: Readings Held

Blink:  Auto-Capture Mode

Pump Speed Limit Reached

P1/P2 Mode:  P1-P2

Test Mode: Shows when 

you are Increasing (InC) or 

Decreasing (dEC) pump speed.

Indicates Whether P1 or P2  

is Shown on Top Display Line

Low Battery