Hydro Logger H1
User manual
FIEDLER
electronics for ecology
77
Follow these steps to update the firmware:
▪
Connect to the unit whose software you want to update with MOST
▪
Read the current parameters from the connected telemetry station
and save them as a parameter file as a backup of the current instru-
ment settings. Similarly, back up archived measurement data.
▪
In MOST, select the "
Firmware
" option in the "
Manufacturing
" menu
▪
Select the desired data file containing the firmware (*.hex) and load
it into the PC by pressing the "Open" button.
▪
After the new firmware transfer is complete, the unit will automati-
cally restart and MOST will display the message "Firmware update
complete".
7.2.
Procedure for replacing the power supply battery
It is advisable to change the battery in the time interval between measurements
when the Hydro Logger is in a resting state and has minimum current consump-
tion (the MEASURE LED is illuminated when a measurement is in progress). If
the battery is changed within 1 minute, the real-time counter will not be dela-
yed and the correct functioning of the instrument will not be impaired. Ne-
vertheless, it is advisable to check the real time and date on the unit's display
after changing the battery and adjust them if necessary. The date and time will
also be compared automatically the next time the HydroLogger H1 communica-
tes with the server.
The power battery is inser-
ted in the holder inside the
device. Therefore, before
replacement, remove the
outermost plastic screw caps
and loosen the 4 corner
screws until the lid of the in-
strument can be removed.
Remove the battery from the
holder and replace it with a
new lithium battery type
LSH20 or equivalent with a
nominal voltage of 3.6V and a capacity of at least 13 Ah. You can order a new
battery from the manufacturer or your Hydro Logger H1 supplier, where you
can also send the used battery for environmentally friendly disposal.
You can disconnect the connecting flat cable between the instrument cover and
the connection plate when replacing the battery. If the instrument is also powe-
red from an external power supply, it is preferable not to disconnect the con-
necting cable, as it guarantees power not only to the time circuit of the Hydro
Logger H1 but also to all other parts of the instrument, including the connected
sensors or GSM module.
When the replacement is complete, be sure to tighten the screws in the lid pro-
perly to prevent moisture from entering the unit, which could condense on the
cold metal parts and damage the Hydro Logger H1's electronic circuitry.
R e p l a c i n g t h e
b a t t e r y
Содержание H1
Страница 1: ...Small telemetry station for water applications H1 H2 Hydro Logger user manual version 1 10 ...
Страница 2: ...Hydro Logger H1 User manual 2 electronics for ecology FIEDLER ...
Страница 80: ...Hydro Logger H1 User manual 80 electronics for ecology FIEDLER Notes ...
Страница 81: ...Hydro Logger H1 User manual FIEDLER electronics for ecology 81 ...