Hydro Logger H1
User manual
FIEDLER
electronics for ecology
43
When using a server, simply save the new parameter file to the server and the
station will download the file at the next data communication.
6.2.
Basic parameters
A group of basic parameters for setting station identification, measurement
frequency, supply voltage control, data archiving and station communication
with sensors.
IDENTIFICATION
The basic parameters include the instrument nameplate, which can store a ma-
ximum of 17 ASCII characters characterizing the instrument. The name tag is
preferably used to visually check the affiliation of an open parameter file to the
connected instrument and can be seen at the beginning of the *.dta data file,
inserted into sent SMS and serves as a station identifier on the server.
Note: the "Device Name" parameter is automatically transferred to the station
name on the server. Therefore, if you need to rename a station on the server,
just change the name tag in the parameters and the next time the station has a
data session with the server, the name of the station on the server will automa-
tically change.
The value of this parameter serves as a station identifier on the server and is
assigned by the server administrator. The
identification number
is stored toge-
ther with the data in one data file and thus uniquely identifies the origin of the
measured data. The parameter can take values in the range 1-65535.
Setting this parameter to a non-zero integer prevents further parameter chan-
ges without knowing the password. The password value can take values from
zero to 9999. A zero value disables the check and allows unlimited overwriting
of parameters.
I n s t r u m e n t
n a m e p l a t e
I d e n t i f i c a t i o n
n u m b e r
P a s s w o r d t o
c h a n g e
p a r a m e t e r s
A password value different
from 0 will block access to
the parameters.
The interval at which the
instrument is switched on
and all inputs are mea-
sured. For each channel,
you can specify its multi-
ples for archiving (then the
average of all measure-
ments for the archiving in-
terval is archived)
Day boundaries for calcu-
lating daily flows.
SMS message archiving
including text, phone num-
ber, date and time.
The analog signal
measurement delay
from power on all-
ows sensors with a
long rise time of a
valid signal from
their connection to
the supply voltage to
be connected
Selecting the size of
the supply voltage
Unap, which is swit-
ched on only during
the
measurement
period
Permanent
swit-
ching on of the
supply voltage for
the OPTO water me-
ter sensor
Содержание H1
Страница 1: ...Small telemetry station for water applications H1 H2 Hydro Logger user manual version 1 10 ...
Страница 2: ...Hydro Logger H1 User manual 2 electronics for ecology FIEDLER ...
Страница 80: ...Hydro Logger H1 User manual 80 electronics for ecology FIEDLER Notes ...
Страница 81: ...Hydro Logger H1 User manual FIEDLER electronics for ecology 81 ...