6.5 Internal Function Table
No.
Description
Parameters / Note
Default =
**
F1
Internal Analogue to Digital
(AD) Value
Press
[Print/M+]
to set offset value to zero when unloaded. Then add load on the platform to observe the
span value of load applied.
•
LCD2 represents the million digits of the AD value.
•
LCD3 represents the AD value up to 999,999.
For example: -
•
LCD2 displays
1
•
LCD3 displays
083690
Then the actual AD value is 1,083,690.
Press
[Zero]
to quit to F1.
F2
All Segment Check & Battery
Voltage
All display segments and indicators will be lit on. Check any segments or indicators are missing.
Press
[Zero]
to quit to F2 or
[Check]
to show current battery voltage.
F3
Capacity, Division & Default
Weight Unit
Display basic metrology characteristics (capacity, division, and weight unit).
•
Capacity x Division of MAX
1
is shown on LCD2. Value displayed = Max
1
+ d
1
.
•
Capacity x Division of W
2
(if set) is shown on LCD3. Value displayed = Max
2
+ d
2
.
F4
Date Format & Date
DD/MM/YY
** YY/MM/DD
MM/DD/YY
•
Press
[MR]
or
[PF/Set]
to change date format and press
[Print/M+]
to confirm.
•
Then press
[Print/M+]
to check current date value.
•
To change date value, press
[Print/M+]
then enter date value and press
[Print/M+]
to confirm.
•
Instrument will then convert and display the month and date value entered into week and day number (0 ~ 6 for Sunday ~ Saturday)
of the year value entered.
•
Select Date Output formats: - 2 formats are available: -
➢
Date = Normal date format as per input above.
➢
Week = WWYYDD format. WW = week number; YY = year value; DD = weekday (Monday = 01… Sunday = 07).
•
Press
[MR]
or
[PF/Set]
to select Date Output Type and press
[Print/M+]
to confirm.
F5
Time
HH:MM:SS in 24-hour format
To adjust time, press
[Print/M+]
, then enter new value and press
[Print/M+]
to save.