53
Appendix D: - System Parameter Inquiry Commands
61
System parameter inquiry commands are used to check system parameter settings. Command format as below: -
a.
Hex code 0D (ASCII code $0D), then
b.
Hex code 0A (LF) (ASCII code $0A), then
c.
Command code listed on below table (all commands are case sensitive), then
d.
Hex code 20 (ASCII code $20).
Command
Code
Description
Number & Description Responded Parameters
Notes: -
a.
If more than one parameter, semi colon separation is inserted
between parameters.
b.
Response from instrument always end up with Hex code 0D 0A
Aa
Calibration weight
unit and application
d1
= calibration weight unit: - 0 = kg; 1 = lb.
d2
= application: - 0 = none; 1 = OIML; 2= NTEP
Ab
Decimal format,
decimal point of kg, g
and lb
d1
= decimal format: - 0 = dot (fixed)
d2
= kg: - 0 = no decimal; 1 = 1 decimal place…… 4 = 4
decimal place
d3
= g: - 0 = no decimal; 1 = 1 decimal place…… 4 = 4 decimal
place; n = not applicable
d4
= lb: - 0 = no decimal; 1 = 1 decimal place…… 4 = 4 decimal
place
Ac
Capacity1 of kg, g, lb
d1
= capacity in kg. Data length = 8 including decimal with
leading space (Hex code 20)
d2
= capacity = g. Data length = 8 including decimal with
leading space (Hex code 20)
d3
= capacity = lb. Data length = 8 including decimal with
leading space (Hex code 20)
Ad
Capacity 2 of kg, g, lb
d1
= mode: - 0 = Off; 1 = dual weighing range; 2 = dual interval
d2
= capacity in kg. Data length = 8 including decimal with
leading space (Hex code 20)
d3
= capacity = g. Data length = 8 including decimal with
leading space (Hex code 20)
d4
= capacity = lb. Data length = 8 including decimal with
leading space (Hex code 20)
Ae
Division 1 of kg, g, lb
d1
= kg: - 0 = 1; 1 = 2; 2 = 5; 3 = 10; 4 = 20; 5 = 50
d2
= g: - 0 = 1; 1 = 2; 2 = 5; 3 = 10; 4 = 20; 5 = 50
d3
= lb: - 0 = 1; 1 = 2; 2 = 5; 3 = 10; 4 = 20; 5 = 50
Af
Division 2 of kg, g, lb
d1
= kg: - 0 = 1; 1 = 2; 2 = 5; 3 = 10; 4 = 20; 5 = 50
d2
= g: - 0 = 1; 1 = 2; 2 = 5; 3 = 10; 4 = 20; 5 = 50
d3
= lb: - 0 = 1; 1 = 2; 2 = 5; 3 = 10; 4 = 20; 5 = 50
Ag
Gravity Factor of
Calibration Place and
Operation Place
d1
= gravity factor of calibration place. Data length = 8 including
decimal with leading space (Hex code 20).
d2
= gravity of location of operation place. Data length = 8
including decimal with leading space (Hex code 20).
Ah
Linearity
Compensation
Function
0 = Off; 1 = On
Ai
ad value of zero
point (offset) value,
weight value of LD1,
d1
= ad value of zero point. Data length = 8 including decimal
with leading space (Hex code 20).
61
Connect the external peripheral which generates system parameter inquiry commands to Comport 2 and set Comport 2 to CMD.