31
12. Static Check Function
37
38
39
40
41
This function is used to compare current weight result with the preset LO and HI Limit. The comparison results
(LO, OK, HI) will then be displayed on the Light Bars with or without buzzer
42
. Targets of Check mode are: -
•
Weighing mode = weight value.
•
Piece Count mode = piece value.
•
Auto Tare Accumulation mode = weight value.
12.1 Set LO& HI Limit
Follow the below steps to set LO and HI Limit.
a.
During desired operation mode, press
[Check]
.
b.
Instrument displays Low followed by a 6-digital value. Enter the LO limit through numeric keys or press
[CE/x10]
to set LO limit to zero then press
[Print/M+]
to save.
c.
Instrument displays HIGH followed by a 6-digital value. Enter the HI limit through numeric keys or press
[CE/x10]
to set LO limit to zero then press
[Print/M+]
to save.
d.
Check Mode is now enabled. The check result is shown by one of the Light Bar
43
.
12.2 Hints for Entering LO and HI Limits
a.
For normal comparison, set both LO and HI Limit.
b.
To check only if result is lower than or equal to LO (result ≤ LO), set HI Limit = 0.
c.
To check only if result is higher than or equal to HI (result ≥ HI), set LO Limit = 0.
d.
To check if result is equal to a specified value, set both HI Limit and LO Limit = the specified value.
12.3 To Cancel Check Function
To cancel check function, press
[CE/x10]
on point
b
and
c
of
12.1
.
12.4 Tri-color Check Result Light Bar
This instrument is equipped with a tri-color light bar. The tri-colors are: -
•
Yellow to represent LO status.
•
Green to represent OK status.
•
Red to represent HI status.
12.4.1 Setting Light Bar working mode
44
5 light bar working modes are available: -
•
Li Bar 0 = Light Bar off.
•
Li Bar 1 = Progress Mode. Under this mode, light bar starts from zero and all the way to current
position.
•
Li Bar 2 =Single Bar Mode. Under this mode, light bar shows only the current result by lighting up
all 8 x LEDs of respective check result.
•
Li Bar 3 = Single Dot Mode. Under this mode, light bar shows only the current result by lighting up
a single LED dot of respective check result. Position of the dot depends on actual result.
37
Check mode will not operate when weight is less than 20d (or 20d
1
for dual weighing range/interval mode).
38
Set also F15 for desired Check buzzer output, light bar working mode and action on negative value
39
When F25 = Mode 1, set also F26 (Near Zero weight value).
40
Check mode does not support peak hold or animal weighing mode.
41
Set F26 to zero.
42
Set F15 for preferred buzzer output configuration.
43
Yellow = Lo; Green = OK; Red = Hi.
44
Light Bar is only visible when Check Function is employed.