5. Initial Setup
There are 2 groups of internal function: -
Group #1: - F1~F32 are accessible without restriction.
Group #2: - F60~F99 are restricted functions which request a password or hardware key to access. These
functions are for dealer and authorized personnel only. Do not change any settings of these functions to
avoid operation errors.
Below paragraphs describe those settings related to F1~F32.
5.1 Internal Functions & Settings
Application parameters can be checked and set through internal function. Set all preferred operation parameters
according to
5.4.
5.2 How to Enter & Select Internal Function
11
Follow the below procedures for internal function setup: -
a.
In weighing function, press
[Set]
.
b.
Instrument displays F1 and is now in internal function mode.
c.
Press
[Func]
and
[Unit]
to access the preferred internal function number.
d.
Quick access to a function number: -
Press
[1]
to go to F10.
Press
[2]
to go to F20.
Press
[3]
to go to F30.
Press
[6]
to go to F60 (for dealer and authorized personnel only).
Press
[8]
to go to F80 (for dealer and authorized personnel only).
Press
[9]
to go to F90 (for dealer and authorized personnel only).
Press
[0]
to go to F1.
5.3 Key Function under Internal Function Mode
Key
Function in Setup & Calibration
[On/Off]
Quit without saving and power off
[Unit]
Go to previous page
[Set]
To enter internal function number F1~F32 during weighing mode
[Print/M+]
Enter, save and return
[Zero]
Quit without saving
[Tare]
Set F1 value being shown to zero and to display the net span gain of additional
load applied
[Func]
Go to next page
[CE/x10]
Clear
[Check]
Quick access to label settings
11
Internal function mode can only be accessed when instrument is in weighing mode.