Fetal Doppler i100 Скачать руководство пользователя страница 1

Fetal Doppler

Operation Manual

MODEL

:

i100

Содержание i100

Страница 1: ...Fetal Doppler Operation Manual MODEL i100 ...

Страница 2: ...T O L i s t e n A n g e l s H e a r t b e a t ...

Страница 3: ...use 4 3 4 The standard configuration 4 Chapter four Product appearance 5 4 1 The host 5 4 2 Display icon 5 4 3 Controlling Keys 6 4 4 Indicator light 6 4 5 probes 6 Chapter Five Basic operation 7 5 1 Used to prepare 7 5 2 To use the battery 7 5 3 The probe using 7 5 4 starting up 7 5 5 Patterns and features 7 5 6 power off 8 5 7 Battery replacement or charging 8 ...

Страница 4: ...Chapter six Maintenance and troubleshooting 9 6 1 maintenance 9 6 2 troubleshooting 9 Chapter seven Warranty and after sales service 11 7 1 The warranty 11 7 2 After sales service 11 ...

Страница 5: ...rior written consent should not be allowed to copy or any part of this manual translated into other languages 2 The meaning of the icon in this product is as follows Prevent electric shock level BF type equipment IPX4 Waterproof level Waterproof splash equipment Pay attention to It means to Mild dangerous The information prompt It is information about the user s connection use or maintenance It is...

Страница 6: ...n to 1 Please read this manual carefully before using this instrument to ensure that you are familiar with control and operation 2 The used battery must be processed according to local regulations 3 If you use rechargeable batteries to ensure the performance and life of rechargeable batteries prior to the first use of the battery power once full of general will need to recharge the battery of cont...

Страница 7: ...s the instrument cannot work when charged 4 If the instrument is not used for a long time please remove the battery in time 5 This instrument is a fetal heart rate examination tool and the results are not the only standard for fetal health 6 This instrument is designed to be a continuous working type and be careful not to be wet by water 7 Do not pasteurize or use electron beam and gamma radiation...

Страница 8: ... probes 1 pcs Function i100 I i100 II i100 III i100 IV Display B W Colour B W Colour Back light Yes Yes Yes Yes Build in speaker Yes Yes Yes Yes Battery indicator Yes Yes Yes Yes Curve display No Yes No Yes Auto power off Yes Yes Yes Yes Alkalinity battery Yes Yes Optional Optional Ni MH battery Optional Optional Yes Yes adapter No No Optional Optional USB charging line No Yes Optional Optional ea...

Страница 9: ...rt 4 2 To display the meaning of the object 4 2 1 i100 I Show the number of placental heart rate and work parameter 4 2 2 i100 LED Display menu heart rate number heart rate curve and working parameters 4 2 3 Show the explanation of the content serial number symbol explain 1 Heart rate symbol 2 MHz Frequency unit insert probe display 3 bpm Fetal heart rate units times minutes 1 2 3 4 5 6 7 10 8 9 ...

Страница 10: ...light 4 4 1Charge indicator When the battery is charged the light is indicated When the battery is full the indicator is green 4 4 2 Working indicator light When the instrument is turned on the indicator is blue 3 5 probes 3 5 1 The basic frequency of the probe is 2 0MHz and 3MHz optional 3 5 2 Probes key is a counter When you feel fetal movement press it once The value on the screen will be accum...

Страница 11: ...te direction of the battery to close the battery storehouse 5 3 Operation of the probes Hold the device in one hand and the other on the top of the probe Remove the head first and then remove the entire probe along the direction of the arrow Pay attention to When turning on do not remove or reload the probe The probe should be removed before turning on and then the probe will be returned after shu...

Страница 12: ...n up or down according to the function of the menu to open or close the current entry 3 seconds without operation exit menu Settings 5 6 Turn off the machine When the machine is turned on press the switch machine key for three seconds then turn off the job light is out Note The machine shuts down automatically when there is no operation in one minute 5 7 Battery replacement or charging When the ba...

Страница 13: ...current insulation including leakage current measurement It is recommended to be tested in accordance with public organization s specifications and inspection procedures 3 The accuracy of the fetal heart rate is controlled by the instrument and cannot be adjusted by the user 4 If the outcome of the fetal heart rate is not credible use other methods such as using a stethoscope to verify or contacti...

Страница 14: ... host 2 Interference of external signals 3 Low battery capacity Solution 1 Keep the probe away from the host 2 Away from the outside signals 3 Replace the battery or charge 6 2 4 Low sensitivity main reason 1 The probes position is incorrect 2 There is no or no coupling enough agent Solution 1 Ensure that the probes is in the right place 2 Apply enough coupling agent ...

Страница 15: ... the warranty scope the whole machine or part The following is not covered under warranty 1 Malfunctions caused by misuse negligence accident and transport 2 Disassemble the modification and repair equipment 3 Change or remove the serial number and the device label 7 2 After sales service If there are any problems in use maintenance technical parameters and instrument failure contact local dealers...

Отзывы: