Section 2: Installation
2.5 Grounding Requirements
0101-8242-0, Rev. C
2-5
CV-6SLX User Manual
2.5.2 System Low-Impedance Grounding
Within the vacuum system, the low-impedance ground is provided by 3”-wide and 1/2”-wide
copper straps. As Figure 2-3 shows, these straps must connect:
•
the grounding stud labeled
RF GND
on the power module’s rear panel (see Figure 2-4) to a
grounding point on the frame of the operator station
•
the operator station’s grounding point to the vacuum cubicle’s central grounding point
•
the grounding stud labeled
RF GND
on the FPS front panel (see Figure 2-5) to the vacuum
cubicle’s main grounding point
•
the source tray to the vacuum cubicle’s main grounding point.
Figure 2-3 System Low Impedance Grounding Requirements
2.5.3 Power Module Grounding
Make connections to the power supply’s grounding lugs as shown in the CV-6SLX Quick Start
Guide (PN 0101-8241-1), following the procedure described below.
Step
Action
1
A roll of 1/2”-wide copper strap (PN 5621-0032-0) is supplied with
the unit. Cut off a length of this strap that will easily extend from
the grounding lugs on the power module’s rear panel to the
operator station’s frame.
Содержание Temescal CV-6SLX
Страница 4: ...CV 6SLX User Manual ii 0101 8242 0 Rev C...
Страница 16: ...1 5 TemEBeam Controller EBC Section 1 Introduction to the CV 6SLX CV 6SLX User Manual 0101 8242 0 Rev C 1 4...
Страница 34: ...2 6 Cable Installation Section 2 Installation CV 6SLX User Manual 0101 8242 0 Rev C 2 18...
Страница 90: ...5 5 FPS Fuse Replacement Procedure Section 5 Troubleshooting CV 6SLX User Manual 5 32 0101 8242 0 Rev C...
Страница 92: ...Appendix A Pinout Table for Rear Panel Diagnostic Port J3 CV 6SLX User Manual A 2 0101 8242 0 Rev C...