© Ferno / 234-3801-00 / August 2021
30
7 - MAINTENANCE
Maintenance
WARNING
Improper maintenance can cause injury. Maintain the
chair only as described in this manual.
Lubricating the track system can result in injury to patient
and/or operators. Never lubricate the tracks or belts.
7.2 Disinfecting/Cleaning Restraints
Ankle Restraint:
Place buckled restraint in a mesh bag and launder in
a washing machine using warm water, a disinfectant soap, and gentle
agitation. Hang the restraint to dry. Do not put it in a dryer.
Chest and Lap Restraints:
Disinfect and clean restraints as directed in
the restraint users’ manual provided. Additional, free users’ manuals can
be obtained from Ferno Customer Relations (page 2).
7.3 Disinfecting/Cleaning the Chair
To disinfect:
Remove restraints and wipe all surfaces of the chair
and panels with disinfectant. Follow the disinfectant manufacturer’s
directions for application method and contact time.
To clean:
Wash panels and all parts of the chair with warm water and
a mild detergent. Rinse with clean water. Dry with a towel or allow to
air-dry.
7.4 Cleaning Tracks and Belts
To ensure predictable belt performance, it is important to keep the track
and belts free of oil, grease, dirt, and debris.
1. Hold the belts away from the track frame and remove any debris.
2. Clean both sides of the track frame, and both belts, with SD-20
™
All-Purpose Cleaner
and a clean cloth. Follow instructions on the
container.
3. Dry track and belts thoroughly.
To order SD-20
™
Cleaner in North America, contact the Spartan
Chemical Company, Maumee, Ohio, USA at 1.800.537.8990 or visit www.
spartanchemical.com on the internet.
7.1 Maintenance Schedule
The chair requires regular maintenance. Set up and follow a
maintenance schedule. The table at right represents minimum intervals
for maintenance.
Keep maintenance records. A sample maintenance record sheet is
provided on page 42.
When using maintenance products, follow the manufacturers’ directions
and read the manufacturers’ material safety data sheets.
NOTICE
Disinfectants and cleaners containing bleach, phenolics, or
iodines can cause damage. Use caution when disinfecting
and cleaning with products containing these chemicals.
Using abrasive cleaning compounds or applicators on the
chair can cause damage. Do not use abrasive materials to
clean the chair or PowerTraxx.
DO NOT PRESSURE WASH
Water under high pressure, or steam, can
penetrate joints, flush away lubricant, and
cause corrosion. Do not pressure wash the
chair or PowerTraxx.
Minimum
Maintenance
Each Use
As Needed
Monthly
Disinfecting
(this page)
•
Cleaning
(this page)
•
Inspection
(page 32)
•
•
7.5 Lubrication-Free Chair and Tracks
The chair and tracks are constructed with lubrication-free materials.
Never lubricate the track, the PowerTraxx, or any other part of the
track system or chair.
Applying lubricant can cause damage to the
chair. Lubricants on the belts or track can cause the chair to perform
unpredictably, resulting in injury to the patient and/or operators.
Do Not Lubricate
Содержание Transcend Chair Series
Страница 39: ... Ferno 234 3801 00 August 2021 39 Parts and Service 1 5 6 8 4 9 2 3 7 ...
Страница 42: ... Ferno 234 3801 00 August 2021 42 Date Maintenance Performed By Maintenance Record MAINTENANCE RECORD ...
Страница 43: ... Ferno 234 3801 00 August 2021 43 NOTES ...
Страница 44: ...Aviation Military Mortuary EMS Rescue 70 Weil Way Wilmington OH 45177 Ferno 234 3801 00 August 2021 ...