Front View and Connections
DP1 = XX.X
Signal
4 5
3
2
1
+
-
A B C D
Offset light P1
(±20 points)
P2 Span light
(±50 points)
P3 Span
DP2 = X.XX
Internal Configuration
Offset P4
Signal
Close Jumpers
4/20 mA
J1 - J4 - J5 - J6
0/20 mA
J1 - J4 - J5 - J6
0/10 Vdc
J3 - J6
0/100 Vdc
J2 - J6
0/450 Vdc
J4 - J11
Note .- J12a always closed. J7,
J10, J12b always open.
J1
Upper Module
seen on solder side
J11
J6
J9 J8 J7
J3
J4
J10
J2
J5
J12a
J12b
Offset Type
Jumpers
Offset Negative J8 Closed
(by default)
J9 Open
Offset Positive J8 Open
J9 Closed
Power
-
+
Manufacturer FEMA ELECTRÓNICA, S.A.
Pol. Ind. Santiga - Altimira 14
E-08210 - Barberà del Vallès - BARCELONA
ESPAÑA - SPAIN
www.fema.es - [email protected]
Series
VIR-30
Models
01,03,04,05,06,08,10,15,20,21,32,51 and 90
The manufacturer declares that the instruments indicated
comply with the directives and rules indicated below.
Directive of electromagnetic compatibility 2004/108/CEE
Directive of low voltage 73/23/CEE
Safety Requirements 61010-1
Equipment “Fixed”, “Permanently connected”
Pollution degree 1, Isolation Double, CAT-II
Emmision requirements
UNE EN 50081-1 (1994)
Immunity requirements
UNE EN 50082-1 (1998)
Barberà del Vallès, April 2010
Daniel Juncà - Quality Manager
CE Declaration of Conformity
INSTALLATION - The power supply circuit must be provided
with a device that permits disconnection if needed. The power
supply type and value, and its connection, are printed on the
labels attached to the instrument. The instrument does not
have internal protection fuse. To meet safety requirements,
the installation of a fuse is needed with the values indicated
below :
230Vac
115Vac
Fuse
50 mA
100 mA
MAINTENANCE -
Before any operation of maitenance,
adjustment, repair or substitution, the instrument must be
unplugged from the power supply. To clean the instrument
use a humid cloth. Do not use abrasive products (solvents,
alcohols, ...) to clean the front filter and plastic parts. Do not
operate the instrument in places with excess of humidity,
smoke, or flammable gas. The instrument is designed to be
mounted on metalic panel.
IN CASE OF FIRE 1- Disconnect the instrument from the
power supply. 2- Give alarm according to local rules. 3- Switch
off air conditioning devices. 4- Attack the fire with carbonic
snow. Do not use water. In closed areas do not use systems
with vaporized liquids.
Prescriptions
Mechanical Dimensions (mm)
69
32,5
Panel cut-out
15
60
5
5 max
72
36