User’s Manual S40-R
User’s Manual S40-R
FEMA ELECTRÓNICA - Page 5
Range and scaling
Pot.<5K
Scaling 0/99.99%=0/99.99%
Alarms 1,2 and 3
Active
Off (not managed)
Type
maximum
Setpoint
1000
Hysteresis
0 counts
Delay
0.0 seconds
Setpoint2
Off
Display
Fixed digits
Off
Average
Off
Steps
Off
Left zeros
Off
Add zero to the right
Off
Memory of maximum -1999
Memory of minimum
9999
Peak & Hold
Off
Tools
Segments
Off
Password
Off
Brightness
3
1.9 Default factory configuration
The instrument has two menus accessible to the user :
“
Configuration Menu”
(key SQ)
“
Information Menu”
(key UP)
The
“Configuration Menu”
allows to change the configuration
of the instrument. Access to the
“Configuration Menu”
can be
password protected with the function “
PASSWORD
”. During
operation with the
“Configuration Menu”
the alarms are kept
“
on-hold
”. When leaving the
“Configuration Menu”
the instru-
ment performs a restart, and new configuration is applied. On
restart of the instrument, also the control output modules are
restarted (relays, analogue outputs, ...).
The
“Information Menu”
is for information only, and it does not
accept changes on the displayed information. To enter the
“In-
formation Menu”
press the “
UP
” button. It is not affected by
the “PASSWORD” function. Leaving the
“Information Menu”
returns to the measuring state of the instrument, without
restart of the unit.
Rollback -
After 30 seconds without interaction from the
operator, the instrument leaves the menu and returns to the
previous working mode. In case of configuration menu, all
changes are discarded.
Button SQ -
Selects the menu entry currently displayed.
When entering a numeric value (for example a setpoint value)
validates the value on display.
Button UP -
Moves vertically on the menu entries. When
entering a numeric value (for example a setpoint value)
modifies the current digit by increasing its value up from 0 to
1, 2, 3, 4, 5, 6, 7, 8, 9.
Button LE -
Leaves the current menu. Pressing LE several
times will leave all menus. When leaving all menus in the
configuration menu, changes will be saved . When entering
a numeric value (for example a setpoint value) it moves from
one digit to the next. Each digit value can then be modified
with the UP button.
1.8 Operating the menus