Sensors
Carbon Dioxide (CO₂)
SENSOR
Infrared sensor, pyroelectric detector
Nominal Range
0-100%
Full scale resolution
0.01% absolute
Accuracy
±1% absolute and ±3% of measured
value
Sampling Time
6 seconds
Calibration
Annually
Lifetime
> 5 years
Oxygen (O₂) Sensor
Electrochemical
Nominal Range
0-100%
Full scale resolution
0.1% absolute
Accuracy
±0.25% absolute and ±2% of
measured value
Sampling Time
6 seconds
Calibration
Bi-annually
Lifetime
12 months
Unpacking the F-920 Check It!
The F-920 arrives with a hard-sided carrying case, two sets of
batteries and a charger, a removable 4 GB SD card, and several
accessory parts. A sampling port with needle is included for
taking non-destructive samples from packaging. The sampling
port is pictured below, connected to the intake.