Fluorescence Light Assembly:

[email protected]

Page 12

Installing the Fluorescence Illuminator:

Place the fluorescence module (2) on the microscope 

body and tighten the set screw.

Place the binocular / trinocular head on top of the 

fluorescence module and tighten the set screw (3).

Insert transformer (1) into socket (2) and then plug 

directly into outlet.

Fluorescence Light Adjustment:

Fluorescence / Brightfield adjustment knob (1) can 

be rotated 180 degrees to switch between 

brightfield and fluorescence.

Rotate the light adjustment knob (2) clockwise to 

make the light brighter.

Содержание RB30

Страница 1: ...RB30 Microscope User s Manual info feinoptic com...

Страница 2: ...info feinoptic com Page 1 Microscope Components...

Страница 3: ...igh temperatures dust or damp environments Use a flat work surface Set the power switch to off O before replacing a bulb or fuse and wait until the lamp is cool The power switch 4 is located on the si...

Страница 4: ...he detergent is removed from the frame with a clean damp cloth prior to drying the surface Do not disassemble the microscope After use cover the microscope with a dust cover and power off the light Ob...

Страница 5: ...Binocular Trinocular Microscope Assembly info feinoptic com Page 4...

Страница 6: ...Fluorescence Microscope Assembly info feinoptic com Page 5...

Страница 7: ...operating the microscope always start using the lowest objective 4x or 10x to get the sample into focus before moving to a higher magnification Eyepiece Assembly Remove the cover from the eyetube 1 In...

Страница 8: ...e brightness and counterclockwise to decrease brightness Using the bulb at a lower brightness extends bulb life Slide Placement Push the specimen holder 1 back Place the slide 2 in the slot with the c...

Страница 9: ...until the image is clear If multiple people are using the microscope note your diopter setting for quick transition from users Interpupillary Distance Adjustment When observing with both eyes adjust t...

Страница 10: ...p with the inscription 4 of the objective being used Using the 100x Immersion Oil Objective Start with the 4x objective and focus on the sample Place a drop of immersion oil 1 on the specimen Rotate t...

Страница 11: ...eyepieces are in focus adjust the focus on the c mount with an allen key to parfocal the camera with the eyepieces Phase Contrast Slider Assembly Insert the slider 1 from left to right with the inscri...

Страница 12: ...move one eyepiece and insert the phase centering telescope Set the phase adjustment ring 2 to correspond with the phase objective being used example PH10x with 10x objective Adjust the centering teles...

Страница 13: ...the binocular trinocular head on top of the fluorescence module and tighten the set screw 3 Insert transformer 1 into socket 2 and then plug directly into outlet Fluorescence Light Adjustment Fluores...

Страница 14: ...r glass thickness 0 17mm Cover slip is on the bottom of the slide Put cover glass face up Immersion oil has dried on objective lens Clean objective lens No oil immersion was used with 100x lens Use im...

Страница 15: ...ched properly Adjust specimen holder Stage is not fastened properly Reattach the stage Power is not turned on Check power cable connection LED light is not inserted properly Check light connection LED...

Страница 16: ...to each 3 Disconnect the wires from the bulb assembly and unscrew the bulb assembly from the bulb compartment 4 Solder the wires to the new bulb assembly in the same location as the old bulb assembly...