Fluorescence Assembly:
Page 14
Fluorescence Illuminator Assembly:
Remove the caps (1) from the illuminator.
Place the illuminator on the microscope body then
slide it toward the light source to make the
epi-illuminator flat with the body.
Tighten the four M5 hex screws in the illuminator
with the spanner and replace the caps (1).
Fluorescence Filter Set Assembly:
Loosen the right side set screw (1) on the turntable
fluorescence illuminator with a M4 inner hex wrench
and pull the front cover (2) out of the dovetail
groove.
The blinker board (5) is installed in the fluorescence
filter group. When using the fluorescence filter
group, first loosen the set screw (6) with the inner
hex wrench and remove the blinker board (5).
Place the diaphragm slice of the fluorescence filter
group (7) which should be assembled upward, and
match it with the dovetail wedge from the front
group (2) and push down. Tighten the set srew (6).
Check the ID (8) on the dovetail interface and insert
the nameplate (4) of the fluorescence filter group
into the interface (3) with the same number in front
of the front cover group (2).
Repeat above steps to assmple other fluorescence
filter groups into the turntable.
Match the dovetail wedge of the front cover group
(2) with the dovetail groove on the turntable
epi-illuminator, and push down. Tighten the set
screw (1).