Step-by-Step Assembly:
Page 11
Stage Adjustment:
When looking through the microscope, move the
stage by rotating the X-Axis adjustment knob (1) and
the Y-Axis adjustment knob (2). The movement
range of the X-Axis and Y-Axis is 80 x 55mm.
X-Axis & Y-Axis Knob Adjustment:
Hold the X-Axis knob (1), tighten the Y-Axis knob (2)
to expose the adjustment knob.
Rotate the X-Axis knob (3) or the Y-Axis knob (4) in
clockwise (the direction of the arrow shown at right)
to reduce tension, or increase the tension by rotating
the knob counterclockwise.
If the tension is too tight, a creaking sound will be
heard from the stage, or the accuracy of the stage
stop will be reduced.
Stage Rail Adjustment:
After extended years of use, the stage rail may
become offset and the movement range may
become shortened. Follow these steps to fix it.
Horizontal: Hold the sample holder, and move the
stage rail left and right until you hit the limit stop.
Vertical: Hold the top surface of the stage and move
the sample holder back and forth until you hit the
limit stop.
Reflected Illumination: Lower the stage bracket. The
microscope can adapt to view a sample of no more
than 35mm, which is useful when observing thick
objects.
Move the stage to the lowest position, then remove
the stage from the microscope.
Loosen the stage bracket lock screw (1) and remove
the stage bracket.
Rotate the coarse focusing knob and raise the focus
board (2) to the position where the limit screw (3)
can be seen from the mirror arm.
Loosen and remove the limit screw (3).
Reinstall the stage bracket and the stage.