System Options:
Heating Stage 1000 °C / Heating Stage 1500 °C
9-48
Water Cooled Temperature Stage Operation Termination
Same as for the Cooling stage (see above).
HS MAINTENANCE
Cleaning
The Heating stage could be polluted after a long term use or when
using highly outgassing samples. To prevent another samples
contamination remove and clean the stainless steel cover plate. The
ceramic paper heater cover should be replaced together. Use Heater
cover 2 (harder paper) always on the top.
The Heat shield isolation should be replaced as well (for HS 1500
only).
Heater breakdown
The HS 1500 assembly lifetime is limited and depends strongly on the
working temperatures.
FIGURE 9-42
1500 °C HEATER LIFETIME
In case of heater failure, check its resistivity, which should be within
the 1 - 5 Ohm range. When the heater breaks down, the 1500 °C
Heating Stage assembly can be ordered.