8-1
8
MAINTENANCE AND TROUBLESHOOTING
This section describes necessary microscope maintenance
procedures that can be carried out by the
FEI Supervisor User
/
FEI
Microscope User
. The user maintenance is at a minimum due to gun
and column design providing the long uptime. Therefore a
complicated maintenance is normally a part of a service contract to be
performed by a qualified service engineer.
At the User level items such as the following can be kept:
•
Cleaning Procedures Overview
•
The Standard Insert
•
Gaseous Detectors
•
Stage Maintenance
•
Refilling the Water Bottle
•
Rotary Pump Maintenance
C a u t i o n !
- Parts that operate in vacuum should be handled carefully using
clean powder-free gloves. Parts not in use should be stored in suitable
containers or packed in aluminium foil.
- the
EDX window
(if present)
is very fragile
and
must be protected
from large pressure oscillations
. It is recommended to remove the
detector before major cleaning activities.
Note:
Gas back fill (N
2
) should be maintained while the specimen chamber
is at ambient pressure. However, to avoid gas waste it is
recommended that the chamber should be left vented no longer than
necessary.
Содержание Quanta FEG 250
Страница 8: ...C vi...
Страница 28: ...System Operation Quanta FEG System States 3 10...
Страница 108: ...Alignments 1 Gun Alignment 6 6...
Страница 114: ...Alignments 154 Water Bottle Venting 6 12 154 Water Bottle Venting...
Страница 152: ...System Options Automatic Aperture System 9 4...
Страница 154: ...System Options Nav Cam 9 6 401 NAV CAM AUTOBRIGHTNESS...
Страница 177: ...System Options Cooling Stage Waterless Cooling Stage 9 29...