FEASA LED ANALYSER LOW LIGHT
FUNCTIONAL VERSION
Set Data Mode
set/putINTGAIN - Set the Intensity GainFactor
Transmit
Receive
Setintgain##xxx
putintgain##xxx
OK
OK
Where:
##
represents the Fiber Number and is a number in the range 01 – 10.
xxx
represents a 3 digit gain factor, default 100.
Description
This command allows the user to adjust the Intensity Gain Factor for each Fiber.
This is useful when it is required to balance all or some of the Fibers to give the same Intensity
when testing similar LED's. The Factory default setting is 100 and the value for each Fiber
can be adjusted from 050 – 200.
These set values are stored permanently in memory and can only be changed by using the
setIntGain
command again.
The put values are stored in RAM memory and must be changed by using the
putIntGain
command each time after Power On/Off.
will display the current stored gain setting.
Example:
Set the Intensity gain for Fiber 1 to 095.
setintgain01095
OK
Example:
Set the Intensity gain for Fiber 1 to 095.
putintgain01095
OK
42