Capture Commands
The Capture commands are used to capture the parameters (color, intensity, xy, uv,
wavelength, cct) of the LED's to be tested and store the results in the memory of the Analyser.
These results can be read out later using the
GET DATA
commands.
Commands are transmitted and received using ASCII characters and are
NOT case-sensitive
.
All commands must be terminated with a
<CR>
or
<LF>
character. All responses from the
Led Analyser are also terminated with <
CR
> <
LF
>
The
Terminal Program
supplied on the CD is used to send/receive commands to/from the
Analyser. This program is also available as a drop down function of the
USER
and
Test
Software programs.
The
Feasa
LED Analyser User
Program is a graphical tool that can be used to send
commands and receive results from the Analyser. It allows one LED to be tested at a time.
This Program also allows a Terminal Window to be opened so that the User can type the
commands directly and send them to the Analyser. The responses from the Analyser can be
observed in the Window.
The
Feasa LED Analyser Test
Program allows the User to test all the LED's together. Pass
and Fail limits can be set and results can be printed and logged.
Alternately, the User may generate a customised Program that sends commands and receives
data through the virtual USB Com Port.