In-Circuit Test Mode
Initialize the LED Analyser for Capture
This function will initialize the LED Analyser and make it ready to capture new LED Test data.
This function should be executed before each capture cycle.
Figure 13 shows the timing for this function.
Figure 13.
The timing sequence is as follows:-
1/
Drive WE_bar (pin14) High and drive OE_bar (pin6) Low. The RyBy_bar (pin18) will be
driven low by the Analyser.
2/
Drive the address bus to 11111 (0x1F) (pins 9-13),This will Initialize the Analyser
3/
Drive WE_bar Low and the Analyser will drive RyBy_bar high for approx 1mS. All
the frequencies (from the previous capture) will now be turned off at this time. After
the time T
rst
= 1mSec, RyBy_bar will be driven low by the Analyser to indicate it is ready
for the next function. The Analyser is now Initialized.
4/
Drive WE_bar High any time after RyBy_bar has gone Low.
24
WE_bar
OE_bar
Addr
0X1F (11111B)
RY_BY
t
rst
Fout
(1ms)