FEASA LED ANALYSER
FUNCTIONAL VERSION
Get Data Mode
getWAVELENGTH##
- Get the Dominant Wavelength
Transmit
Receive
getwavelength##
getwavelengthall
XXX
Where:
##
represents the Fiber Number and is a number in the range 1 – 20.
all
represents the 20 Fibers in a Feasa 20F, or 3 Fibers in a 3F.
xxx
represents the dominant wavelength of the LED in nanometers.
Description
This command is used to get the value of the Dominant Wavelength for the LED under the
Fiber number
##
.
This command should be preceded by the
capture C
command to ensure valid LED data is
stored in the memory of the LED Analyser. The
C1-C5
commands are available but are not as
accurate. The data for each LED can be read out one-at-the-time, in any order.
Alternatively use the
getwavelengthall
command to read back the LED data for all 20
leds in one command.
Example:
The PC transmits
getwavelength01
to the LED Analyser to instruct it to send the stored
dominant wavelength for the LED positioned under Fiber No 1. The LED Analyser will return a
string
xxx
to the PC.
getwavelength01
513
31