
FEASA LED ANALYSER
FUNCTIONAL VERSION
Get Data Mode
getxy## - Return the xy Chromaticity values
Transmit
Receive
getxy##
getxyall
0.xxxx 0.yyyy
Where:
##
represents the Fiber Number and is a number in the range 01 – 20.
all
represents the 20 Fibers in a Feasa 20F, or 3 Fibers in a 3F.
0.xxxx
represents the x Chromaticity value
0.yyyy
represents the y Chromaticity value
Description
This command is used to return the xy Chromaticity value for the LED under the Fiber number
##. This command is used for testing White LED's.
This command should be preceded by a
capture
command to ensure valid LED data is stored
in the memory of the LED Analyser. The data for each LED can be read out one-at-the-time,
in any order. Note: it is important for the PC to use 01, 02, 03, etc to indicate Fibers 1, 2
and 3, etc.
Alternatively use the
getxyall
command to read back the LED data for all 20 leds in
one command.
Example:
The PC transmits
getxy01
to the LED Analyser to instruct it to send the stored xy Chromaticity
data for the LED positioned under Fiber No 1. The LED Analyser will return a string
0.xxxx
0.yyyy
to the PC.
getxy01
0.6461 0.3436
27
Содержание 10-F
Страница 11: ...FEASA LED ANALYSER FUNCTIONAL VERSION Figure 3b Intensity for the LED Analyser in LINEAR Mode Back to Index 11...
Страница 126: ...FEASA LED ANALYSER FUNCTIONAL VERSION User Software Back to Index 126...
Страница 127: ...FEASA LED ANALYSER FUNCTIONAL VERSION Test Software Back to Index 127...
Страница 128: ...FEASA LED ANALYSER FUNCTIONAL VERSION User Calibration Back to Index 128...
Страница 129: ...FEASA LED ANALYSER FUNCTIONAL VERSION Programming Back to Index 129...