Capture Mode
CAPTUREMULTI - Multiple Capture Command
Transmit
Receive
setrange@@#
putrange@@#
getrange@@
capturemulti
OK
OK
1 to 5
OK
Where:
@@
represents the fiber number in the range 01 – 20.
#
represents the capture range 1 - 5
Description
The
capturemulti
or
cm
command will capture the LED data for the specified fiber at the
specified range.
The
setrange@@#
or
putrange@@#
commands will set the required range for each declared
individual fiber. The
getrange@@
will return the range the user has enables for this fiber.
Example:
The DUT has 5 leds with 5 very different intensity values. You can now specify fiber 1 to range
Ultra (C5), Fiber 2 to range Super (C4), Fiber 3 to range High (C3), Fiber 4 to range Medium
(C2) and Fiber 5 to range Low (C1). Do one
capturemulti
or
CM
command to get your 5
readings.
Putrange015
(Fiber 1, Range Ultra / 5)
putrange024
(Fiber 2, Range Super / 4)
putrange033
(Fiber 3, Range High / 3)
putrange042
(Fiber 4, Range Medium / 2)
putrange051
(Fiber 5, Range Low / 1)
capturemulti
23
Содержание 10-F
Страница 11: ...FEASA LED ANALYSER FUNCTIONAL VERSION Figure 3b Intensity for the LED Analyser in LINEAR Mode Back to Index 11...
Страница 126: ...FEASA LED ANALYSER FUNCTIONAL VERSION User Software Back to Index 126...
Страница 127: ...FEASA LED ANALYSER FUNCTIONAL VERSION Test Software Back to Index 127...
Страница 128: ...FEASA LED ANALYSER FUNCTIONAL VERSION User Calibration Back to Index 128...
Страница 129: ...FEASA LED ANALYSER FUNCTIONAL VERSION Programming Back to Index 129...