DazScope
manual
Chapter 3 : Software overview
Figure 3.6: Content of the spectrometer configuration panel.
Integration time:
integration time of the spectrometer expressed in the units indicated on
the right on the numeric control.
Average:
number of spectra to acquire for a given acoustic wave. The displayed spectrum is
the average of all acquired spectra.
Lambda Min:
lowest wavelength to acquire (expressed in the units specified on the right
side of the numeric control)
Lambda Max:
highest wavelength to acquire (expressed in the units specified on the right
side of the numeric control)
Open From File...:
loads the parameters from an XML file.
Save To File... :
save the current parameters to an XML file.
Reset To Default:
resets the parameters to their default values.
Make Default:
sets the current parameters as default parameters.
3.3.2
Configuring the ChirpScan algorithm
Double-click on the
ChirpScan
item of the
Algorithm
listbox. The configuration panel of the
ChirpScan panel opens (figure
During the chirp scan successive waveforms are defined and sent into the Dazzler and the
corresponding SHG spectra are recorded and displayed. All waveforms are strictly identical, but
for the chirp value which is scanned over a given interval. This interval is defined as a variation
FASTLITE-9 september 2008