FaroArm User Guide
January 2007
39
Chapter 2: CAM2 Measure Devices Menu
Sphere Method
The Sphere calibration is performed
using any size precision sphere or tooling
ball. The sphere should be at least 10mm
(3/8") diameter or larger.
You will digitize 15 individual points in
specific locations around the sphere by
pressing the FRONT button for each.
C
AUTION
: The ball probe
must
remain in
contact with the sphere while digitizing
points. Even one or two poorly digitized
points significantly affects the optimization process, which then has an
effect on the accuracy of the FaroArm.
•
Digitize five points around
the top of the sphere with the
probe pointing in position
#1.
•
Digitize five points around
the front of the sphere with
the probe pointing in
position #2.
Position #1
Position #2
Position #3
1
2
3
4
5
1
2
3
4
5
08M46E00_FaroArmUSB.book Page 39 Monday, January 15, 2007 1:12 PM
Содержание FaroArm P08
Страница 1: ...FAROARM USER GUIDE JANUARY 2007 08M46E00_FaroArmUSB book Page 1 Monday January 15 2007 1 12 PM...
Страница 2: ...08M46E00_FaroArmUSB book Page 2 Monday January 15 2007 1 12 PM...
Страница 4: ...08M46E00_FaroArmUSB book Page 4 Monday January 15 2007 1 12 PM...
Страница 48: ...08M46E00_FaroArmUSB book Page 42 Monday January 15 2007 1 12 PM...
Страница 54: ...08M46E00_FaroArmUSB book Page 4 Monday January 15 2007 1 12 PM...
Страница 64: ...08M46E00_FaroArmUSB book Page 10 Monday January 15 2007 1 12 PM...
Страница 70: ...08M46E00_FaroArmUSB book Page 6 Monday January 15 2007 1 12 PM...
Страница 76: ...D 6 08M46E00_FaroArmUSB book Page 6 Monday January 15 2007 1 12 PM...
Страница 80: ...08M46E00_FaroArmUSB book Page 4 Monday January 15 2007 1 12 PM...
Страница 81: ...08M46E00_FaroArmUSB book Page 1 Monday January 15 2007 1 12 PM...