FaroArm User Guide
January 2007
34
Chapter 2: CAM2 Measure Devices Menu
Single Hole Method - Guidance
The Single Hole calibration is performed using the FARO probe
calibration cone. If the G
UIDANCE
check box is selected for the current
probe, use the following steps.
All of the points in this method will be collected by holding down the
FRONT button. The FaroArm will collect points as fast as possible
(“scanning”) until you release the FRONT button. You will collect 200
points in three different positions.
Figure 2-3 Single Hole Method
1
Place the ball probe in the
cone.
2
With the probe seated in
the cone, move the handle
down until the shaft of the
probe is parallel with the
top of the cone.
Position #1
Position #2
Position #3
08M46E00_FaroArmUSB.book Page 34 Monday, January 15, 2007 1:12 PM
Содержание FaroArm P08
Страница 1: ...FAROARM USER GUIDE JANUARY 2007 08M46E00_FaroArmUSB book Page 1 Monday January 15 2007 1 12 PM...
Страница 2: ...08M46E00_FaroArmUSB book Page 2 Monday January 15 2007 1 12 PM...
Страница 4: ...08M46E00_FaroArmUSB book Page 4 Monday January 15 2007 1 12 PM...
Страница 48: ...08M46E00_FaroArmUSB book Page 42 Monday January 15 2007 1 12 PM...
Страница 54: ...08M46E00_FaroArmUSB book Page 4 Monday January 15 2007 1 12 PM...
Страница 64: ...08M46E00_FaroArmUSB book Page 10 Monday January 15 2007 1 12 PM...
Страница 70: ...08M46E00_FaroArmUSB book Page 6 Monday January 15 2007 1 12 PM...
Страница 76: ...D 6 08M46E00_FaroArmUSB book Page 6 Monday January 15 2007 1 12 PM...
Страница 80: ...08M46E00_FaroArmUSB book Page 4 Monday January 15 2007 1 12 PM...
Страница 81: ...08M46E00_FaroArmUSB book Page 1 Monday January 15 2007 1 12 PM...