Chapter 1: Introduction to the FaroArm
®
Plane Calibration
Calibrate the FARO Laser Line Probe. You will digitize the FARO
Calibration Plate, once with the ball probe, and then three times with the
Laser Line Probe.
1 Digitize the White area of the Calibration Plate with the ball probe.
• Touch the white surface with the ball probe and press the FRONT
button.
• Digitize eight points on the white surface.
N
OTE
: Digitize each point near an edge of the surface to avoid
marring the middle of the surface.
• Pull the probe away from the surface and press the BACK button.
The flatness of the white surface is calculated and the results are
displayed in the dialog box. If the results
Fail
, click the RETRY button
to redigitize the white surface. Click OK to continue.
N
OTE
: Avoid touching both lenses. Clean the top and bottom lenses
with the cloth from the Laser Line Probe case. Dirt and grease on
either lens can cause poor results.
8
7
6
5
4
3
2
1
08M46E00_FaroArmUSB.book Page 37 Tuesday, February 17, 2009 1:42 PM
Содержание FaroArm Fusion
Страница 2: ...08M46E00_FaroArmUSB book Page 2 Tuesday February 17 2009 1 42 PM ...
Страница 4: ...08M46E00_FaroArmUSB book Page 4 Tuesday February 17 2009 1 42 PM ...
Страница 8: ...08M46E00_FaroArmUSB book Page iv Tuesday February 17 2009 1 42 PM ...
Страница 100: ...08M46E00_FaroArmUSB book Page 92 Tuesday February 17 2009 1 42 PM ...
Страница 113: ...B 9 08M46E00_FaroArmUSB book Page 9 Tuesday February 17 2009 1 42 PM ...
Страница 114: ...08M46E00_FaroArmUSB book Page 10 Tuesday February 17 2009 1 42 PM ...
Страница 120: ...08M46E00_FaroArmUSB book Page 6 Tuesday February 17 2009 1 42 PM ...
Страница 129: ...08M46E00_FaroArmUSB book Page 1 Tuesday February 17 2009 1 42 PM ...