Continuous
250 V RMS, Measurement Category II
Withstand
1,390 V, verified by a 5 s dielectric withstand test
Channel-to-earth ground
Continuous
250 V RMS, Measurement Category II
Withstand
2,300 V, verified by a 5 s dielectric withstand test
Explosive atmospheres
Channel-to-channel
60 V DC, Measurement Category I
Channel-to-earth ground
60 V DC, Measurement Category I
NI-9229 with BNC Safety Voltages
Connect only voltages that are within the following limits:
Isolation
Channel-to-channel
Continuous
60 V DC, Measurement Category I
Withstand
1,000 V, verified by a 5 s dielectric withstand test
Channel-to-earth ground
Continuous
60 V DC, Measurement Category I
Withstand
1,000 V, verified by a 5 s dielectric withstand test
Safety Guidelines for Hazardous Voltages
If hazardous voltages are connected to the product, take the following precautions.
A hazardous voltage is a voltage greater than:
■
30 V RMS, 42.4 V peak, or 60 V DC in DRY LOCATIONS
■
16 V RMS, 22.6 V peak, or 35 V DC in WET LOCATIONS