7
Test Setup
From this menu the user can enter or reset a temperature ramp and soak schedule.
The user can also configure the hesitation squeeze and gel strength parameters on
equipped instruments. This section explains how to use the software to view and
print the test results.
Temperature Control (Pressure When Available)
7.1
The Program Temperature screen is shown below:
Enter desired set point values by touching the SET POINT parameter box and
entering the values on the keypad. Time to set point is selected by touching the
TIME parameter box and entering a time in minutes on the touch pad. The final
segment should always be a SOAK segment where the temperature equals the final
test temperature and the time equals SOAK. This guarantees the instrument will
maintain final temperature for the remainder of a test. After a profile has been
entered, press the SAVE button. At this point a graph of the desired temperature
Touch to program temperature. A
similar box appears on pressure
controlled units for pressure ramp.
Change any additional test