
Cement Curing Autoclave User Manual
100072577
Revision H
41
5.7
Pressurize the chamber
5.7.1
Connect the pressure source to the manual valve.
5.7.2
Open the manual valve at least one (1) and not more than two (2)
full turns.
5.7.3
Operate the pressure source to increase the pressure inside the
curing chamber to the desired initial value as required for the test. See
API RP 10B.
5.7.4
Close the manual valve to retain the chamber pressure.
5.7.5
If desired, the pressure source may be disconnected.
Figure 5-21
Set initial chamber pressure
5.8
Running a test
5.8.1
Adjust the set point of the temperature controller to the desired
maximum temperature. See API RP 10B for settings.
5.8.2
Program Temperature Control for a Test
5.8.2.1
From the home display, press until you reach
’ProG
LiSt’
. Press.
5.8.2.2
Use the arrow keys to select the number of the program to
be changed. Press.
5.8.2.3
For each program setting, use
or
to change the
displayed value. After releasing the button, the display will blink to show the
controller has accepted the new value.
Содержание 100072577
Страница 1: ...Cement Curing Autoclave User Manual 100072577 Revision H ...
Страница 10: ...Cement Curing Autoclave User Manual 100072577 Revision H 3 Figure 1 2 Cement Curing Autoclave back view ...
Страница 15: ...Cement Curing Autoclave User Manual 100072577 Revision H 8 Figure 3 3 Cement Mold Figure 3 4 Cement Mold Clamp ...
Страница 18: ...Cement Curing Autoclave User Manual 100072577 Revision H 11 ...
Страница 36: ...Cement Curing Autoclave User Manual 100072577 Revision H 29 Figure 5 2 Clamps hold mold halves together ...
Страница 40: ...Cement Curing Autoclave User Manual 100072577 Revision H 33 Figure 5 7 Filling mold Figure 5 8 Leveling cement ...
Страница 63: ...Cement Curing Autoclave User Manual 100072577 Revision H 56 Figure 9 4 Assembly Cement Curing Autoclave ...