Fahrenheit  Digital  Clinical  Thermometer

PLEASE READ CAREFULLY BEFORE USING

ORAL USE

1. Close mouth for about 5 minutes before measurement.

2. Place the thermometer tip under the tongue on a heat pocket (

3

) as shown in

Fig. 7.

3. Hold the thermometer in the same spot for about 40 seconds.

4. Readings will continue to increase and the 

o

F symbol will flash during mea-

surement (See Fig.8).

5. When peak temperature is reached, the 

o

F symbol will stop flashing (See Fig.

9) and

a)

When the temperature is higher than 99.9 

o

F the beeper will sound “beep-

beep-beep” ten times to indicate a high temperature.

b)

When the temperature is lower than or equal to 99.9 

o

F the beeper will

sound  “beep”  ten  times.

6. After measurement, record the temperature, then turn off the unit to save the

battery life. If you forget to turn it off, the unit will power down automatically
after approximately 8 minutes.

NOTE

 : THE MOUTH MUST REMAIN CLOSED DURING THE MEASUREMENT TO ENSURE

ACCURATE READINGS.

RECTAL USE

Rectal temperature should only be taken when oral and axillary temperatures are inconvenient. It is
recommended that axillary temperatures be taken on infants. LUBRICATE WITH LUBRICATING
JELLY AND INSERT THE PROBE NO MORE THAN 1/2 INCH INTO RECTUM. Hold thermometer
in place until you hear the beep signal and the 

o

F sign stops flashing (about 40 seconds.)

AXILLARY USE

Dry armpit before measurement. Place the tip of the probe under arm so that the tip is touching the skin.
This area is called axilla. Hold the thermometer so that it is parallel to the patient’s body and fold the
arm over the chest. This ensures that the tip of the probe is well covered and not influenced by air. Hold
thermometer in place until you hear the beep signal and the 

o

F sign stops flashing (about 2 minutes)

BATTERY REPLACEMENT

Replace battery with an alkaline LR41 or equivalent type whenever the low battery indicator “ + -  ”
flashes.

1. Remove screw located on the back of the thermometer. Press

down on the battery cover and push in the direction show
by the arrow in the illustration.

2. Remove battery with a non-metallic probe.

3. Insert new battery. NOTE : The positive (+) side must face

outward.

4. Replace the cover.

5. Dispose off used battery properly.

DISPOSABLE  PROBE  COVER  USE.

1. Insert the thermometer into probe cover. Take care not to push the

thermometer tip through the end of the probe cover.

2. Holding the thermometer and tab of the probe cover as shown, pull

away the covering from longer side.

3. Tilt thermometer to remove cover from shorter side.

4. Dispose off the probe cover after taking temperature.

NOTE : For rectal use, after applying probe cover lubricate thermometer

tip with a water soluble jelly.

LIFETIME GUARANTEE

Your new Faichney Dual Scale Digital Clinical Thermometer is guaranteed to be free from manufacturer’s
defects for life under normal use.

However, if for any reason (other than abuse or normal battery replacement) your thermometer does
not function properly we will either repair or replace it free of charge for life.  This warranty is valid
only if unit returned for repair is properly packed to avoid damage in shipping.  All parts are covered by
the warranty, except batteries and probe covers, where applicable.

Please check the battery before returning it. This is usually the source of the problem, should repair be
needed, ship freight paid to :

Conforms to ASTM   E 1112

2122308  Rev-02

Fig.7

1.

2.

3.

BODY TEMPERATURE

Normal body temperature can vary from  97

o

F to 99 

o

F . The generally accepted “Normal” temperature

is 98.6 

o

F. Usually your temperature is lower in the morning than in the afternoon. Normal temperature

varies according to the location of the temperature reading. While the oral temperature follows the
guidelines  previously  described,  a  rectal  temperature  is  generally  1.0 

o

F  higher.  Conversely,  an

axillary (under the arm) temperature will be 1.0 

o

F lower.

GENERAL CARE AND USE

BEFORE USE PLEASE NOTE

1. The thermometer should be held during use. The patient should be seated or lying down.

2. TEMPERATURES SHOULD BE TAKEN UNDER ADULT SUPERVISION.

SPECIAL PRECAUTIONS

3. Avoid exposure to temperature extremes, dust, excessive humidity and direct sunlight.

4. Do not soak the thermometer in water or other liquids. Keep parts other than the probe away

from moisture.

5. Do not attempt to sterilize with heat or by boiling.

6. Do not bend or drop the unit. It is not shock proof.

7. Do not allow the unit to come into contact with any thinners or chemical solvents.

8. Do not attempt to disassemble the unit except to replace the battery.

9. Do not bite the main body or sensor unit.

DIRECTIONS  FOR  CLEANING

Clean the Thermometer probe before and after use by wiping it with soap and water. Rinse with
70%  Isoprophyl Alcohol,  then  rinse  thoroughly  in  cold  water.

CAUTION

NEVER

 SUBMERGE THIS THERMOMETER IN ANY LIQUID.

NEVER

 ATTEMPT TO STERILIZE WITH HEAT, BY BOILING OR IN A GAS OR STEAM

AUTOCLAVE.

SPECIFICATIONS

DISPLAY  RESOLUTION

1/10 DEGREE FOR 

O

F

MEASUREMENT RANGE

89.6 

O

F to 109.4 

O

F

ACCURACY

±0.3 

O

F (89.6 

O

F to < 95.0 

O

F )

±0.2 

O

F (95.0 

O

F to < 104.0 

O

F )

±0.3 

O

F (>104.0 

O

F to 109.4 

O

F )

MEASUREMENT METHOD

PEAK HOLD SYSTEM

BATTERY

ONE 1.5V DC ALKALINE MANGANESE BUTTON

SIZE

LR41 TYPE OR EQUIVALENT

BATTERY LIFE

UPTO 3 YEARS UNDER NORMAL USE

AUTOMATIC SHUT-OFF

AFTER APPROXIMATELY 8 MINUTES

OPERATING TEMPERATURE

16 

O

C to 40 

O

C (61

O

F to 104 

O

F)

RELATIVE HUMIDITY

15% to 95% NON CONDENSING

STORAGE TEMPERATURE

-20 

O

C to 50 

O

C (– 4 

O

F to 122 

O

F)

RELATIVE HUMIDITY

15% to 95% NON CONDENSING

OPERATING  INSTRUCTIONS

1. Ensure that the thermometer has been turned off before taking temperature.

2. To  start  the  unit  -  Press  the  blue  colour  ON/OFF  button  momentarily

once.

3. The following appear in quick sequence.

(a) This is the LCD display test (See Fig. 1). It appears for approximately

2 seconds and indicates all the segments of the LCD are functioning
properly.

(b) If the previous reading is between 89.6 

o

F ~ 109.4 

o

F, the  unit will

display the last time measured temperature value for about 3 seconds
with symbol M (Fig.2.) If the last time measure temperature value is
less  than  89.6 

o

F  the  unit  will  display  Lo 

o

F  (Fig.  3)  for  about  3

seconds without symbol M and if the last meassured reading is greater
than 109.4

o

F, the unit will display Hi (Fig. 4) without symbol M.

(c) The unit goes through the internal self cheching test procedure. If the

unit is good it will display 98.6 

o

F (Fig. 5) for about one second.

(d) When  “Lo”  appears  with  the  flashing 

o

F  sign  (Fig.6)  the  unit  is

ready to take temperature.

Digital Display Window

Battery Cap

Sensor

On / Off Switch

E

M

F

M

Fig. 1

Fig. 2

Fig. 3

Fig. 4

Fig. 5

Fig. 6

F

F

F

F

Faichney  Medical  Company

17517,  Fabrica  Way  Suite  G
Cerritos  CA  90703,
USA

Faichney Fahrenheit Thermometer

Reorder  No.  6112