FTG Series Programming Manual
_________________________________________________________________
19
3.3. OUTPut command
OUTPut[:STATe]
Turn on/off output, also used for turn on/off test functin output.
Command syntax: OUTPut[:STATe] <bool>
Parameter: 0 | 1 | OFF | ON
Example: OUTP ON
Query syntax: OUTPut?
Returned parameter: <NR1>
OUTPut:FUNCtion
Switch power supply test function.
Command syntax: OUTPut:FUNCtion <NR1>
Parameter: 0 | 1 | 2 | VI | SEQ | CP
Example: OUTP:FUNC VI
Query syntax: OUTPut:FUNCtion?
Returned parameter: <NR1>
OUTPut:PROTect:VOLTage
Set power supply overvoltage protection threshold.
Command syntax: OUTPut:PROTect:VOLTage <NRf>
Parameter: MIN
~
MAX
Unit: V (Volt)
Example: OUTP:PROT:VOLT 10
Query syntax: OUTPut:PROTect:VOLTage?
Returned parameter: <NRf>[Unit=V]
OUTPut:PROTect:CURRent
Set power supply overcurrent protection threshold.
Command syntax: OUTPut:PROTect:CURRent <NRf>
Parameter: MIN
~
MAX
Unit: A (Ampere)