3.2. CONFigure command
CONFigure:FOLD:BACK
Set foldback protection mode, when power supply switches output mode between CC
and CV, the output will be turned off.
Command syntax: CONFigure:FOLD:BACK <NR1>
Parameter: 0 | 1 | 2 | OFF | CV2CC | CC2CV
Example: CONF:FOLD:BACK 1
Query syntax: CONFigure:FOLD:BACK?
Returned parameter: <NR1>
CONFigure:FOLD:TIME
Set foldback protection delay time.
Command syntax: CONFigure:FOLD:TIME <NRf>
Parameter: 0.1
~
600.0
Unit: s (Second)
Example: CONF:FOLD:TIME 1
Query syntax: CONFigure:FOLD:TIME?
Returned parameter: <NRf>[Unit=s]
CONFigure:APG:MODE
Set analog programming mode.
Command syntax: CONFigure:APG:MODE <NR1>
Parameter: 0 | 1 | 2 |3 | 4 | OFF | U | I | U&I | P
Unit: None
Example: CONF:APG:MODE 1
Query syntax: CONFigure:APG:MODE?
Returned parameter: <NR1>
CONFigure:APG:VOLTage
Set analog programming reference voltage.
Command syntax: CONFigure:APG:VOLTage <NR1>
Parameter: 0 | 1 | REF5 | REF10