Annex
Page 18/
20
10.3
Safety instructions
Level Gauge CONDURIX Ex …
Edition: 04.2011
I
Range of application
The intrinsically safe equipment CONDURIX Ex … is used for the continuous measurement of liquid
levels. The level sensor operates only in electrically conductive liquids (conductivity
≥
1
μS/cm). If the
level sensor is inserted into a tank with non
-
conducting wall, the sensor must be equipped with a
counter electrode, e. g. the CONDURIX Ex MA …
II
Standards
See EC-
Type Examination Certificate.
III
Instructions for safe …
III.a
… use
The approval applies to the following types or device versions:
CONDURIX
Ex
…
Mono
CONDURIX
Ex
…
DU
CONDURIX
Ex
…
MA
CONDURIX
Ex
E HY
CONDURIX
Ex
E … V
CONDURIX
Ex
…
extern
CONDURIX
Ex
…
extern
Steck
CONDURIX
Ex
…
M12
All level sensors CONDURIX Ex … can be produced with plastic insulation (e. g., PEEK, PTFE or cera
m-
ic with O ring seal).
In order to vary the height the version CONDURIX Ex E …
V is provided. The process connection is
done via a cutting ring fitting.
With limited mounting space the electronics may be installed in an external enclosure (CONDURIX Ex
… external). The connection to the sensor can be done via a fixed cable or via a connector (e. g.,
LEMO).
The level sensor CONDURIX
Ex
…
HART has in addition to the current signal a digital communication
capability supported by the HART protocol. This allows a highly flexible programming and operation
of the level sensor.
III.b
… mounting
Screw
-
in unit:
Seal the thread of the screw
-
in unit with a suitable sealing material, screw it into the existing
socket and tighten it.
In the case of installation with a cutting
ring fitting, the position of the sensor can no longer be
altered after the union nut has been tightened.
Flange:
The probe tube is permanently fixed to the flange, which means that the installation length
cannot be altered. Seal the flange with a suitable sealing and fix it with the flange bolts or nuts.
If the filling level sensor is supplied without process connection, the installer is responsible for co
m-
pliance with the EX requirements.