background image

1. 

Check all bolts to ensure they are tight. 

Remove covers as necessary for inspection.

2. 

Check arm to mast pivot bolt 

by removing plastic cap to ensure nut is tight  

     and cotter pin is in place. If plastic cap is missing order replacement.

3. 

Foot platform assembly pins need to checked

 to ensure  

    they have not pushed in and still engage into base.

 
4.

 Inspect shin pad for tears or rips.

4.

 Safety tabs need to be checked 

to make sure they are installed correctly,  

    not missing or torn.
 

PIN ON FOOT
PLATFORM

ARM TO MAST

Classic Stand

®

 

Safety

 & 

Maintenance Checklist

PO Box 89, 710 E. Main. Clarinda, IA 51632  

I

  

1-800-627-8940 

I

  

[email protected]

EZ Way, Inc.

The Classic Stand® requires a minimum of servicing to keep it in good working order.  Nevertheless, it is important 
that certain basic checks be periodically made by maintenance staff to ensure on-going safety throughout the life of the 
device.

The manufacturer suggests that the following components and operating points be scheduled for inspection at intervals 
not greater than six months.  Any detected deficiency must be rectified before the stand is put back into service.

1

2

3

4

5

BASE                        PIVOT                 REAR WHEEL        FRONT WHEEL  