7) There should be no gas in the pressure tube filled with liquid.
8) When using the spacer liquid, the liquid in the pressure tubes
on both sides should be the same.
9) When using the cleaning agent, the connection of the cleaning agent
should be close to the pressure tapping of the process pipeline. The
length and diameter of the pipes through which the cleaning agent
passes should be the same. The cleaning agent should be prevented
from passing through the transmitter.
2.3 Electrical Installation
Wiring diagram of system:
Figure 2-6
(Note 1: Please refer to the usage methods of distributor and safety barrier
if the users require the distributor or safety barrier according to the site and
design requirements.)
It is recommended to wire the explosion-proof cable to the terminal, the