Integral Measuring Concept IMC-8D(A)-Biogas2
Article-No.: 430106 (430107)
ExTox Gasmess-Systeme GmbH
●
Max-Planck-Straße 15 a
●
59423 Unna (Germany)
●
Tel. +49 2303 33 247-0
●
Fax +49 2303 33 247-10
Internet: www.extox.de
●
mail to: [email protected]
430106_430107_IMC-8D_A-Biogas2_e_2011-09-20.doc,21/02/2012Seite 1 von 2
Description
Integral
Measuring Concept for monitoring of biogas systems:
Sampling and conditioning of measured gas, transmitter (such as CH
4
,
CO
2
, O
2
, H
2
S) and evaluation combined in one compact wall mounted
housing. Process of measurement especially designed for biogas applica-
tion.
At the same time possibility of continuous monitoring of ambient air
(such as CH
4
, H
2
S).
▪
Eight 4-20 mA-Outputs (only 430107: IMC-8DA-Biogas2)
Features
Transmitter
▪
Number: 1 to 8
▪
Allocation of transmitter inputs to analysis including sampling and condi-
tioning of measured gas as well as monitoring of ambient air freely se-
lectable
▪
Selection out of ExTox-Series Sens(-I) and ExSens(-I)
Signal Processing
▪
Control Unit ET-8D or ET-8DA for 8 transmitter inputs, incl. 20 freely
configurable relay outputs and serial interface; additionally control of
sampling and conditioning of measured gas as well as evaluation of
status messages.
▪
Software Extension Biogas-Analysis : discontinuous measurement in
batch mode with configurable intervals for measuring / flushing / air; in-
dication of the latest measured value during flushing and air phases;
pressure compensation; control of overload protection for transmitters
based on the principle electrochemical cell
▪
Continuous monitoring of ambient air
Sampling of Measured Gas
▪
Gas suction pump and electronic flow rate monitoring
▪
Maximum length of sample line
≥
50 m
▪
Condensate trap incl. hose pump
▪
Magnetic valve to change from measured gas to flushing air and to con-
trol the batch measurement
▪
Overload protection for hydrogen sulphide transmitter
▪
Hosing: PE/PP
▪
Dust filter
▪
Flame arrestor
¯
IIG IIB3 (inlet of measured gas)
Connections
At the bottom of the housing for 4/6-Hose (inner / outer Ø: 4/6 mm)
▪
3 glands for measured gas inlet, flushing air and gas outlet
▪
1 gland for condensate outlet
Operation Temperature
+5 °C to +40 °C
Pressure at sampling point
-100 hPa to +100 hPa (relative to ambient)
Change-over of
measuring points
(Option)
Sampling and
conditioning of
measured gas
Monitoring of ambient air
Sensors