background image

USER

 

GUIDE

 

 
 

 

Coating

 

Thickness

 

Tester

 

 

 

Model

 

CG206

 

 

 

 

Содержание CG206

Страница 1: ...USER GUIDE Coating Thickness Tester Model CG206 ...

Страница 2: ...ferrous metals Intuitive menu based programming Memory storage of up to 1500 readings in 30 groups 50 readings per group Delete single readings or all groups easily Auto Power OFF with disable feature Data transfer to PC via USB User Zero Calibration ability Applications Corrosion protection Paint Shops and Electroplaters Chemical Automobile Shipbuilding and Aircraft industries Laboratory Workshop...

Страница 3: ...tor 6 Back Key 7 Down arrow 8 Sensor Note Battery compartment on rear of meter Display Description 1 Data record memory group number 2 Record Number in current memory group 3 Battery status icon 4 Measurement units µm or mils 5 DEL Delete Press the back button to delete the previous reading 6 Metal type FE ferrous NFE non ferrous 7 Automatic Auto metal detect mode ...

Страница 4: ...a professional application The heavier metal substrate is the ferrous magnetic zero substrate Fe and the lighter metal substrate Al is the non ferrous non magnetic zero substrate In Auto mode the meter automatically senses ferrous and non ferrous substrates 1 Place a reference film 250µm for example on the ferrous substrate FE 2 Power the meter on by pressing the power button 3 Place the spring lo...

Страница 5: ...gs in each group Group 0 has no memory MEASURE MODE NFE FE Auto Non Ferrous Ferrous or Auto Automatic Metal selection SET Unit settings µm Micrometers mil Mils mm 2 54 100 Backlight Increase decrease backlight using up down arrows Auto Power OFF Enable Allows Auto Power OFF Disable Defeats Auto Power OFF feature Contrast Increase decrease contrast using up down arrows Info Shows firmware version n...

Страница 6: ... LEFT button SELECT to make a selection AUTO mode the meter automatically recognizes the metal that is being measured Ferrous FE Mode the Magnetic induction measurement mode is used Non Ferrous NFE Mode the eddy current measurement mode is used 5 Press the RIGHT button to return to the normal operating mode SET 1 Press the LEFT button SEL to access the menu 2 Use the UP DOWN arrows to scroll to SE...

Страница 7: ...el are displayed Press the right button to exit to the SET menu MEASURE VIEW The Measure View menu allows for scrolling through the readings in all of the groups Readings can be viewed or deleted in this mode 1 Press the LEFT button SEL to access the main menu 2 Use the DOWN ARROW button to scroll down to MEASURE VIEW 3 Press LEFT SELECT to open the MEASURE VIEW parameter 4 Use the UP DOWN ARROW b...

Страница 8: ...n data scroll to DEL Zero of FE or DEL Zero of NFE and then press the LEFT button SELECT This executes the Zero delete function and returns the meter to the SET menu 6 Press the RIGHT button to return to the normal operating mode Reference Film Accuracy Verification 1 The CG206 meter comes with a set of reference films to check the measurement accuracy of the meter 2 The base materials are Ferrous...

Страница 9: ...n the LCD s lower level 6 Press the sensor against the FE or NFE zero reference substrate and wait for the beep 7 Press the LEFT button OK to accept the calibration value or press the RIGHT button CANCEL to abort the calibration 8 If the meter did not read 0 0 you can try again or Set the Zero reference by pressing and holding the ZERO button until the meter beeps 9 When the Zero calibration is co...

Страница 10: ...00 for the 30 groups To delete all groups or an individual group use the MEASURE VIEW mode as explained in the programming menu section of this User Guide In GROUP mode if the storage capacity is exceeded The meter will display 50 in the upper right corner and will beep twice SOFTWARE This meter is capable of storing data and can download that data to a PC through the supplied Windows compatible s...

Страница 11: ... meter is not to be used for 60 days or more remove the batteries and store them separately Battery Replacement Installation instructions 1 Open the rear battery door 2 Replace install the two 1 5V AAA batteries observing proper polarity 3 Secure the battery compartment Never dispose of used batteries or rechargeable batteries in household waste As consumers users are legally required to take used...

Страница 12: ...er of Min area 275 6mils 7mm 196 9mils 5mm Basic critical thickness 19 69mils 0 5mm 11 81mils 0 3mm Industrial standards Conforms to GB T 4956 1985 GB T 4957 1985 JB T 8393 1996 JJG 889 95 and JJG 818 93 Operating Temperature 0 C 40 C 32 F 104 F Operating Relative Humidity R H 20 90 Relative Humidity Power 2 x AAA batteries with Auto Power OFF after 3 minutes Dimensions 120 x 62 x 32 mm 4 7 x 2 4 ...

Отзывы: