Performing Basic Maintenance Operations
58
XTA-50 User Guide
5.
Follow the instructions displayed on screen to update the firmware version.
Once the update is finished, the XTA-50 starts normally.
6.
Remove the USB device.
6.2
Cleaning Optical Connectors
Subject
To optimize the performance of the instrument and prevent loss of optical power, you
must verify that optical connectors are clean every time you connect a fiber.
The XTA-50 optical connectors are mounted on a drawer to ease the cleaning of internal
connectors.
Before Starting
Make sure you have the following material:
•
Optical grade cleaning cotton swabs
•
Canned air
•
Isopropyl alcohol
•
Fiberscope or similar if available
•
Lint-free tissue or cleaning cartridges
Use only high quality cleaning supplies that are non-abrasive and leave no residue
Procedure
1.
Turn the XTA-50 off (see section
) and unplug the power
supply cord from the wall socket.
2.
On the front panel, make sure the protective caps of the connectors are in place.
3.
On the front panel, use your fingers to unscrew the two screws of the connector
drawer.
Once unscrewed from the front panel, the two screws stay attached to the drawer.
4.
Gently pull the drawer out of the front panel (no more than 70 mm) so that fiber ends
are made visible, as illustrated in the following figure.
Figure 26: Optical Connectors – Cleaning
5.
At the rear of the drawer plate, remove one connector end from the plate:
6.
Gently clean the connector end, with the following instructions:
Important
To reduce the need for cleaning, immediately replace protective caps on the
optical connectors when not in use.
Содержание XTM-50 O-band
Страница 1: ...www EXFO com XTA 50 Optical Tunable Filter User Guide XTA 50_UG_2 1v2 3...
Страница 2: ......
Страница 18: ...Installing and Connecting the XTA 50 18 XTA 50 User Guide...
Страница 22: ...Turning on off the XTA 50 22 XTA 50 User Guide...
Страница 62: ...Performing Basic Maintenance Operations 62 XTA 50 User Guide...
Страница 68: ...List of Figures 68 XTA 50 User Guide...