The following procedure explains how to perform wavelength monitoring of the T100S-
HP using a wavemeter.
Procedure
1.
Configure the wavemeter remote interface with the following parameters:
•
Remote: [On]
•
RS-232: [On]
•
Baud: [9600]
•
Handshaking: [None]
2.
Connect the T100S-HP DB9 serial port to the wavemeter serial port.
3.
As the tunable laser acts as the controller device for the wavemeter, press the
2nd
+
Remote
function keys at the T100S-HP front-panel and select:
RS-232 IN: [OFF]
4.
Start the Feedback function by pressing the
2nd
+
mW/dBm
keys.
5.
Enter the wavelength setting at the T100S-HP front-panel.
•
The T100S-HP executes the wavelength monitoring scheme (see
)
so that the output signal wavelength closely matches the wavelength setting.
The message
Moving…
is displayed at the front-panel indicator until the output
wavelength has reached the target value and has stabilized within wavemeter
accuracy.
To compensate for possible disturbance such as accidental vibrations, shocks,
temperature drifts, and possible malfunction of the wavemeter, the T100S-HP
performs the wavelength monitoring scheme with a 4
-
second periodicity.
•
The T100S-HP front-panel display shows the tunable laser operating wavelength,
alternating with the
l FEEDBACK:ON
message to indicate that the "active
wavelength monitoring" is currently on.
If the active wavelength monitoring does not work properly, the
λ
FEEDBACK:ERR
message appears, alternating with the tunable laser operating wavelength.
6.
To turn off the active wavelength monitoring, press the
2nd
+
mW/dBm
again.
The message
FEEDBACK:OFF
displays once then goes off.
Содержание T100S-HP Essential
Страница 1: ...www EXFO com T100S HP High Power Tunable Laser User Guide T100SHP_UM_6 06v1 2...
Страница 2: ......
Страница 10: ...Important Safety Information for Your Product 10 T100S HP User Guide...
Страница 20: ...Product Presentation 20 T100S HP User Guide...
Страница 24: ...Installing and Connecting the Product 24 T100S HP User Guide...
Страница 42: ...Enabling Active Wavelength Monitoring 42 T100S HP User Guide...
Страница 50: ...Using the Auxiliary Inputs and Outputs 50 T100S HP User Guide...
Страница 54: ...Verifying Performances 54 T100S HP User Guide...
Страница 58: ...Performing Basic Maintenance Operations 58 T100S HP User Guide...
Страница 64: ...List of Tables 64 T100S HP User Guide...
Страница 66: ...List of Figures 66 T100S HP User Guide...
Страница 70: ...Index 70 T100S HP User Guide...