Test Setup - Test Configurator, Timer, and System
Power Blazer
167
EtherSAM - Global
Burst Test
check box when selected (cleared by default) verifies
that the expected burst size can be transmitted at maximum burst
rate with minimal loss.
CBS
(Committed Burst Size) check box when selected (see
on page 285), verifies the performance of a committed
burst size at
CIR
’s average
TX
rate.
EBS
(Excess Burst Size) check box when selected (see
on page 285), verifies the performance of an excess
burst size at
CIR+EIR’s
average
TX
rate.
The maximum Jitter, Latency, Frame Loss, and throughput are
measured. For
CBS
, the Jitter, Latency, and Frame Loss are
compared to the SLA thresholds to declare a pass/fail verdict. For
EBS
, the throughput is compared to the SLA thresholds to declare
a pass/fail verdict.
The burst test procedure is generated for each enabled service.
Service Performance Test
check box when selected (default) verifies
that the
SLA
parameters (see
are met
over time by running multiple services simultaneously. The maximum
Jitter, Latency, Frame Loss, and average throughput are measured and
compared to the configured thresholds to declare pass/fail verdicts.
The
Service Performance Test
is only performed for services that have
their
CIR
check box selected.
Subtest Duration
allows setting the duration time, in HH:MM:SS
format, for the
Service Performance Test
(default is 10 minutes).
Global Test Duration Estimate
indicates the total estimate duration of
the test.
Содержание Power Blazer 88000 Series
Страница 1: ...User Guide 88000 Series Power Blazer HIGH SPEED MULTISERVICE TEST MODULE ...
Страница 12: ......
Страница 16: ......
Страница 22: ......
Страница 28: ......
Страница 38: ......
Страница 58: ...Test Setup Test Applications 46 88000 Series OTN BERT ...
Страница 64: ...Test Setup Test Applications 52 88000 Series SONET SDH BERT For SDH BERT ...
Страница 69: ...Test Setup Test Applications Power Blazer 57 SONET SDH DSn PDH BERT For SDH ...
Страница 550: ......
Страница 670: ......
Страница 674: ......
Страница 686: ......
Страница 707: ...Glossary Power Blazer 695 10G Ethernet Client The following figure presents a typical network application ...
Страница 721: ...Glossary Power Blazer 709 Ethernet Cables Straight Through Cable 1000 Mbit s Crossover Cable 1000 Mbit s ...
Страница 765: ...Glossary Power Blazer 753 SONET DSn SDH PDH The GUI Grid indicates the TUG 2 x and TUG 3 x values ...