Test Setup - Test Configurator, Timer, and System
190
MAX-800 Series
Streams - Profile
Total TX Rate
indicates the percentage of the total line utilization
which is the sum of all TX rate enabled streams.
Note:
The Individual stream can be enabled/disabled even when the test is
started and running. The streams can be enabled one after the other, up to
16, as long as the maximum rate is not reached. For example, if the first
stream is using the full rate available, then no other stream can be enabled.
However, if the first enabled stream uses half rate, then at least another
stream can be enabled using up to half rate. Thus, to enable a second
stream, first set the TX rate value within the non-used rate, then enable it. A
stream cannot be enabled if its MAC address is not valid (it can be either
not resolved or wrongly entered).
Link Capacity
indicates the total rate available for traffic generation.
QoS Metrics
Note:
QoS Metrics settings apply to all streams.
Global Pass/Fail Verdict
check box when selected (default) enables
the pass/fail verdict for all streams.
Global Thresholds Type
button
Throughput
allows selecting if the verdict is based on the
Current
Throughput
(default) or
Average Throughput
.
Frame Loss
allows selecting if the verdict is based on a frame loss
Count
(default) or
Rate
.
Out-of-Sequence
allows selecting if the verdict is based on an
Out-of-Sequence
Count
(default) or
Rate
.
Содержание MAX-800 Series
Страница 1: ...www EXFO com User Guide MAX 800 Series MaxTester Application ETHERNET AND TRANSPORT TESTER...
Страница 16: ......
Страница 18: ......
Страница 26: ......
Страница 43: ...Test Setup Test Applications MaxTester 35 SONET SDH BERT For SDH BERT...
Страница 48: ...Test Setup Test Applications 40 MAX 800 Series SONET SDH DSn PDH BERT For SDH...
Страница 62: ......
Страница 208: ......
Страница 414: ......
Страница 426: ......
Страница 446: ...Glossary 438 MAX 800 Series Ethernet Cables Straight Through Cable 1000 Mbit s Crossover Cable 1000 Mbit s...
Страница 473: ...Glossary MaxTester 465 SONET DSn SDH PDH The GUI Grid indicates the TUG 2 x and TUG 3 x values...
Страница 476: ......
Страница 492: ...Index 484 MAX 800 Series Z Z0 344 Z1 346 Z2 349 Z3 353 Z4 353 Z5 353 Z6 355 Z7 356 Zero Code Suppression 169...
Страница 494: ...0 5 1 5 48 5 0 176 W d y K y K D...