Test Setup - Test Configurator, Timer, and System
144
MAX-800 Series
RFC 2544 - Subtests
Accuracy
is the accuracy measurement in percentage of the line rate
(
%
),
Mbit/s
, or in
Gbit/s
. The accuracy is not based on the configured
Maximum Rate
but on the Ethernet line rate:
Note:
For Dual Test Set using different rates on local and remote, the accuracy on
the remote will be set using the same percentage of its line rate as the local.
For example, if the local interface is 1GE with an accuracy of 0.1 Gbit/s
which correspond to an accuracy of 10% and the remote interface is 10GE,
the 10% is applied to its line rate corresponding to an accuracy of 1 Gbit/s.
Acceptable Errors
represents the number of acceptable errors for the
test:
0
(default) to
10
.
Validations
represents the number of times the result should be
validated:
1
(default) to
50
times.
Interface
Speed
Max. Rate
%
Mbit/s
Gbit/s
10 Mbit/s
0.01 to 10.0 (default 1.0) 0.01 to 1.0 (default 0.10)
Not Applicable
100 Mbit/s 0.01 to 10.0 (default 1.0) 0.1 to 10.0 (default 1.0)
Not Applicable
1000 Mbit/s 0.01 to 10.0 (default 1.0) 1 to 100.0 (default 10)
0.001 to 0.100 (default 0.010)
10G LAN
0.01 to 10.0 (default 1.0) 10.0 to 1000.0 (default 100)
0.01 to 1.00 (default 0.10)
10G WAN 0.01 to 10.0 (default 1.0) 10.0 to 1000.0 (default 100.00)
0.01 to 1.00 (default 0.1)
100G
0.01 to 10.0 (default 1.0) 100.0 to 10000.0 (default 1000.0) 0.10 to 10.00 (default 1.0000)
Содержание MAX-800 Series
Страница 1: ...www EXFO com User Guide MAX 800 Series MaxTester Application ETHERNET AND TRANSPORT TESTER...
Страница 16: ......
Страница 18: ......
Страница 26: ......
Страница 43: ...Test Setup Test Applications MaxTester 35 SONET SDH BERT For SDH BERT...
Страница 48: ...Test Setup Test Applications 40 MAX 800 Series SONET SDH DSn PDH BERT For SDH...
Страница 62: ......
Страница 208: ......
Страница 414: ......
Страница 426: ......
Страница 446: ...Glossary 438 MAX 800 Series Ethernet Cables Straight Through Cable 1000 Mbit s Crossover Cable 1000 Mbit s...
Страница 473: ...Glossary MaxTester 465 SONET DSn SDH PDH The GUI Grid indicates the TUG 2 x and TUG 3 x values...
Страница 476: ......
Страница 492: ...Index 484 MAX 800 Series Z Z0 344 Z1 346 Z2 349 Z3 353 Z4 353 Z5 353 Z6 355 Z7 356 Zero Code Suppression 169...
Страница 494: ...0 5 1 5 48 5 0 176 W d y K y K D...