High Performance Power Meter
53
7
Recording Power Signal
Variations
The Min./Max. function allows you to record the extremes of a varying
power signal when performing a continuous acquisition. For example,
it could be used to determine the stability of a light source over time or to
measure the polarization-dependent loss (PDL) of a passive component
when combined with a polarization state controller.
The function is started manually and can be stopped manually or
automatically using the timer function. Minimum (Min.) and maximum
(Max.) values can be recorded and displayed in any measurement unit
(dB, dBm, W, or W/W; see
Selecting the Measurement Unit
In logarithmic scales (dB and dBm), the difference between the maximum
and minimum values is expressed as Max.
Min. In linear scales (W and
W/W), the difference is expressed as a ratio (Min./Max.).
To use the Min./Max. function:
1.
Select the
Continuous
acquisition rate (see