Optical Spectrum Analyzer
437
A Technical Specifications
I
MPORTANT
The following technical specifications can change without notice.
The information presented in this section is provided as a reference
only. To obtain this product’s most recent technical specifications,
visit the EXFO Web site at
www.exfo.com
.
SPECIFICATIONS
a
Spectral Measurement
FTB-5240S and FTB- 5240S-P
FTB-5240BP
Wavelength range (nm)
1250 to 1650
1250 to 1650
Wavelength uncertainty (nm)
b
±0.05
±0.03
±0.01
c, d
±0.01
c, d
Reference Internal
e
Internal
e
Resolution bandwidth (FWHM)
(nm)
f
0.065
b, d
0.033
b, d
Wavelength linearity (nm)
±0.01
b, d
±0.01
b, d
Wavelength repeatability 2
m
(nm)
±0.003
g
±0.002
g
Power Measurements
FTB-5240S and FTB-5240S-P
FTB-5240BP
HPW Option
Dynamic range (dBm) (per channel)
b
–80
h
to +18
–80
h
to +18
–70
h
to +23
Maximum total safe power (dBm)
+23
+23
+29
Absolute power uncertainty (dB)
±0.5
i
±0.5
i
±0.5
i
Power repeatability 2
m
(dB)
d, g
±0.05
±0.04
j
±0.05
j
Optical Measurements
FTB-5240S and FTB-5240S-P
FTB-5240BP
HPW Option
Optical rejection ratio at 1550 nm (dB)
at 0.2 nm (25 GHz)
35 (40 typical)
45 (50 typical)
35 (40 typical)
at 0.4 nm (50 GHz)
45 (50 typical)
50 (55 typical)
45 (50 typical)
Channel spacing
50 to 200 GHz
CWDM
PDL at 1550 nm (dB)
±0.08
d
±0.06
d
ORL (dB)
*
40
*
40
Measurement time
(s)
d, j
<1 (with the FTB-500 platform)
<1 (with the FTB-500 platform)
(includes scanning, analysis and display)
In-Band OSNR Measurements
d, k
FTB-5240S-P
only
FTB-5240BP
OSNR dynamic range (dB)
>35
l
>35
l
OSNR measurement uncertainty (dB)
±0.5
m
±0.5
m
Repeatability (dB)
±0.2
n
±0.2
n
Data signals
Up to 100 Gbit/s
o
Up to 100 Gbit/s
o
Measurement time (s)
d, j
<6 (eight scans with the FTB-500 platform)
<6 (eight scans with the FTB-500 platform)
(includes scanning, analysis and display)
<75 (eight scans with the FTB-200 platform)
Notes
a.
All specifications are for a temperature of 23 °C ± 2 °C with
an FC/UPC connector unless otherwise specified, after
warm-up.
b. From 1520 to 1610 nm.
c. After user calibration in the same test session within 10 nm
from each calibration point.
d. Typical.
e. Integrated and wavelength-independent self-adjustment.
f. Full width at half maximum.
g. Over one minute in continuous acquisition mode.
h. With averaging.
i. At 1550 nm, –10 dBm input.
j. 45 nm span, full resolution, 20 peak analysis.
k. In-band OSNR measurement performed with 64 scans.
l. For an optical noise level of > –60 dBm.
m. With PMD
)
15 ps and no crosstalk, uncertainty specification
is valid for OSNR
)
25 dB. With PMD
)
15 ps and crosstalk,
uncertainty specification is valid for OSNR
)
20 dB.
n. Repeatability specification is valid for OSNR
)
25 dB.
o. Except for pol-mux and fast polarization scrambled signals.
GENERAL SPECIFICATIONS
Temperature
operating
0 °C to 40 °C
(32 °F to 104 °F)
storage
—20 °C to 50 °C
(—4 °F to 120 °F)
Relative humidity
0 % to 95 % non-condensing
Battery life (hours)
5 (with the FTB-500 platform)
Connectors
EI (EXFO UPC Universal Interface)
EA (EXFO APC Universal Interface)
Size (H x W x D)
FTB-5240S module
96 mm x 51 mm x 260 mm
(3
3
/
4
in x 2 in x 10
1
/
4
in)
FTB-5240BP module
96 mm x 76 mm x 260 mm
(3
3
/
4
in x 3 in x 10
1
/
4
in)
Weight
FTB-5240S module
1.5 kg
(3.3 lb)
FTB-5240BP module
1.7 kg
(3.8 lb)
LASER SAFETY
LASER SAFETY
21 CFR 1040.10 and IEC 60825-1
CLASS 1 LASER PRODUCT
Содержание FTB-5240BP
Страница 1: ...User Guide Optical Spectrum Analyzer for FTB 500 FTB 5240S 5240BP...
Страница 8: ......
Страница 14: ......
Страница 54: ...Setting Up the Instrument in WDM Mode 46 FTB 5240S 5240BP Defining Preferences 2 Select the Display tab...
Страница 110: ......
Страница 120: ...Setting Up the Instrument in Drift Mode 112 FTB 5240S 5240BP Defining Preferences 2 Select the Display tab...
Страница 181: ...Setting Up the Instrument in DFB Mode Optical Spectrum Analyzer 173 Defining Preferences 2 Select the Display tab...
Страница 190: ......
Страница 225: ...Setting Up the Instrument in EDFA Mode Optical Spectrum Analyzer 217 Defining Preferences 2 Select the Display tab...
Страница 240: ...Setting Up the Instrument in EDFA Mode 232 FTB 5240S 5240BP Setting Up EDFA Analysis Parameters 2 Select the General tab...
Страница 282: ...Testing DWDM Systems 274 FTB 5240S 5240BP Starting a Measurement Drift Mode DFB Mode...
Страница 283: ...Testing DWDM Systems Optical Spectrum Analyzer 275 Starting a Measurement Spectral Transmittance Mode EDFA Mode...
Страница 306: ...Managing Results 298 FTB 5240S 5240BP Managing WDM Test Results 3 Select the Include in calculation option...
Страница 372: ...Managing Results 364 FTB 5240S 5240BP Managing Spectral Transmittance Test Results 2 Press Global Parameters...
Страница 391: ...Managing Results Optical Spectrum Analyzer 383 Managing EDFA Test Results...
Страница 392: ...Managing Results 384 FTB 5240S 5240BP Managing EDFA Test Results 3 Select the Include in calculation option...
Страница 394: ...Managing Results 386 FTB 5240S 5240BP Managing EDFA Test Results...
Страница 407: ...Managing Results Optical Spectrum Analyzer 399 Managing Trace Files 2 Press New...
Страница 414: ...Managing Results 406 FTB 5240S 5240BP Managing Trace Files...
Страница 422: ...Managing Results 414 FTB 5240S 5240BP Generating Reports A sample WDM txt report is shown below...
Страница 423: ...Managing Results Optical Spectrum Analyzer 415 Generating Reports A sample Drift html report is shown below...
Страница 424: ...A sample Drift txt report is shown below...
Страница 425: ...Managing Results Optical Spectrum Analyzer 417 Generating Reports A sample DFB html report is shown below...
Страница 426: ...Managing Results 418 FTB 5240S 5240BP Generating Reports A sample Spectral Transmittance html report is shown below...
Страница 427: ...Managing Results Optical Spectrum Analyzer 419 Generating Reports A sample EDFA html report is shown below...
Страница 428: ......
Страница 434: ......
Страница 437: ...Troubleshooting Optical Spectrum Analyzer 429 About 2 Press About...
Страница 444: ......
Страница 446: ......
Страница 530: ......